Your browser doesn't support javascript.
loading
Sub-pixel localisation of passive micro-coil fiducial markers in interventional MRI.
Rea, Marc; McRobbie, Donald; Elhawary, Haytham; Tse, Zion T H; Lamperth, Michael; Young, Ian.
Afiliação
  • Rea M; Radiological Sciences Unit (2N), Charing Cross Hospital, London, W6 8RF, UK. marc.rea@imperial.ac.uk
MAGMA ; 22(2): 71-6, 2009 Apr.
Article em En | MEDLINE | ID: mdl-18802730
ABSTRACT

OBJECTIVE:

Electromechanical devices enable increased accuracy in surgical procedures, and the recent development of MRI-compatible mechatronics permits the use of MRI for real-time image guidance. Integrated imaging of resonant micro-coil fiducials provides an accurate method of tracking devices in a scanner with increased flexibility compared to gradient tracking. Here we report on the ability of ten different image-processing algorithms to track micro-coil fiducials with sub-pixel accuracy. MATERIALS AND

METHODS:

Five algorithms maximum pixel, barycentric weighting, linear interpolation, quadratic fitting and Gaussian fitting were applied both directly to the pixel intensity matrix and to the cross-correlation matrix obtained by 2D convolution with a reference image.

RESULTS:

Using images of a 3 mm fiducial marker and a pixel size of 1.1 mm, intensity linear interpolation, which calculates the position of the fiducial centre by interpolating the pixel data to find the fiducial edges, was found to give the best performance for minimal computing power; a maximum error of 0.22 mm was observed in fiducial localisation for displacements up to 40 mm. The inherent standard deviation of fiducial localisation was 0.04 mm.

CONCLUSION:

This work enables greater accuracy to be achieved in passive fiducial tracking.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Interpretação de Imagem Assistida por Computador / Aumento da Imagem / Imagem por Ressonância Magnética Intervencionista / Magnetismo Idioma: En Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Reino Unido

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Interpretação de Imagem Assistida por Computador / Aumento da Imagem / Imagem por Ressonância Magnética Intervencionista / Magnetismo Idioma: En Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Reino Unido