Your browser doesn't support javascript.
loading
Physical and mechanical properties of a thermomechanically treated NiTi wire used in the manufacture of rotary endodontic instruments.
Pereira, E S J; Peixoto, I F C; Viana, A C D; Oliveira, I I; Gonzalez, B M; Buono, V T L; Bahia, M G A.
Afiliação
  • Pereira ES; Department of Restorative Dentistry, Faculty of Dentistry, Universidade Federal de Minas Gerais, Belo Horizonte-MG, Brazil.
Int Endod J ; 45(5): 469-74, 2012 May.
Article em En | MEDLINE | ID: mdl-22188417
ABSTRACT

AIM:

To compare physical and mechanical properties of one conventional and one thermomechanically treated nickel-titanium (NiTi) wire used to manufacture rotary endodontic instruments.

METHODOLOGY:

Two NiTi wires 1.0 mm in diameter were characterized; one of them, C-wire (CW), was processed in the conventional manner, and the other, termed M-Wire (MW), received an additional heat treatment according to the manufacturer. Chemical composition was determined by energy-dispersive X-ray spectroscopy, phase constitution by XRD and the transformation temperatures by DSC. Tensile loading/unloading tests and Vickers microhardness measurements were performed to assess the mechanical behaviour. Data were analysed using analysis of variance (α = 0.05).

RESULTS:

The two wires showed approximately the same chemical composition, close to the 1 1 atomic ratio, and the ß-phase was the predominant phase present. B19' martensite and the R-phase were found in MW, in agreement with the higher transformation temperatures found in this wire compared with CW, whose transformation temperatures were below room temperature. Average Vickers microhardness values were similar for MW and CW (P = 0.91). The stress at the transformation plateau in the tensile load-unload curves was lower and more uniform in the M-Wire, which also showed the smallest stress hysteresis and apparent elastic modulus.

CONCLUSIONS:

The M-Wire had physical and mechanical properties that can render endodontic instruments more flexible and fatigue resistant than those made with conventionally processed NiTi wires.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Titânio / Preparo de Canal Radicular / Ligas Dentárias / Níquel Idioma: En Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Brasil

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Titânio / Preparo de Canal Radicular / Ligas Dentárias / Níquel Idioma: En Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Brasil