Graphene oxide-based electrochemical label-free detection of glycoproteins down to aM level using a lectin biosensor.
Analyst
; 141(14): 4278-82, 2016 Jul 21.
Article
em En
| MEDLINE
| ID: mdl-27277703
A label-free ultrasensitive impedimetric biosensor with lectin immobilised on graphene oxide (GO) for the detection of glycoproteins from 1 aM is shown here. This is the first time a functional lectin biosensor with lectin directly immobilised on a graphene-based interface without any polymer modifier has been described. The study also shows that hydrophilic oxidative debris present on GO has a beneficial effect on the sensitivity of (8.46 ± 0.20)% per decade for the lectin biosensor compared to the sensitivity of (4.52 ± 0.23)% per decade for the lectin biosensor built up from GO with the oxidative debris washed out.
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Glicoproteínas
/
Técnicas Biossensoriais
/
Técnicas Eletroquímicas
/
Grafite
/
Lectinas
Idioma:
En
Ano de publicação:
2016
Tipo de documento:
Article
País de afiliação:
Eslováquia