Importance of a moderate plate-to-bone distance for the functioning of the far cortical locking system.
Med Eng Phys
; 56: 48-53, 2018 06.
Article
em En
| MEDLINE
| ID: mdl-29685856
The far cortical locking (FCL) system, a novel bridge-plating technique, aims to deliver controlled and symmetric interfragmentary motion for a potential uniform callus distribution. However, clinical data for the practical use of this system are limited. The current study investigated the biomechanical effect of a locking plate/far cortical locking construct on a simulated comminuted diaphyseal fracture of the synthetic bones at different distance between the plate and the bone. Biomechanical in vitro experiments were performed using composite sawbones as bone models. A 10-mm osteotomy gap was created and bridged with FCL constructs to determine the construct stiffness, strength, and interfragmentary movement under axial compression, which comprised one of three methods: locking plates applied flush to bone, at 2â¯mm, or at 4â¯mm from the bone. The plate applied flush to the bone exhibited higher stiffness than those at 2â¯mm and 4â¯mm plate elevation. A homogeneous interfragmentary motion at the near and far cortices was observed for the plate at 2â¯mm, whereas a relatively large movement was observed at the far cortex for the plate applied at 4â¯mm. A plate-to-bone distance of 2â¯mm had the advantages of reducing axial stiffness and providing nearly parallel interfragmentary motion. The plate flush to the bone prohibits the dynamic function of the far cortical locking mechanism, and the 4-mm offset was too unstable for fracture healing.
Palavras-chave
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Placas Ósseas
/
Fixação Interna de Fraturas
Idioma:
En
Ano de publicação:
2018
Tipo de documento:
Article
País de afiliação:
Taiwan