Your browser doesn't support javascript.
loading
Noninvasive assessment of steatosis and viability of cold-stored human liver grafts by MRI.
Young, Liam A J; Ceresa, Carlo D L; Mózes, Ferenc E; Ellis, Jane; Valkovic, Ladislav; Colling, Richard; Coussios, Constantin-C; Friend, Peter J; Rodgers, Christopher T.
Afiliação
  • Young LAJ; Oxford Centre for Clinical Magnetic Resonance Research, Radcliffe Department of Medicine, University of Oxford, Oxford, United Kingdom.
  • Ceresa CDL; Nuffield Department of Surgical Sciences, University of Oxford, Oxford, United Kingdom.
  • Mózes FE; Oxford Centre for Clinical Magnetic Resonance Research, Radcliffe Department of Medicine, University of Oxford, Oxford, United Kingdom.
  • Ellis J; Oxford Centre for Clinical Magnetic Resonance Research, Radcliffe Department of Medicine, University of Oxford, Oxford, United Kingdom.
  • Valkovic L; Oxford Centre for Clinical Magnetic Resonance Research, Radcliffe Department of Medicine, University of Oxford, Oxford, United Kingdom.
  • Colling R; Department of Imaging Methods, Institute of Measurement Science, Slovak Academy of Sciences, Bratislava, Slovakia.
  • Coussios CC; Nuffield Department of Surgical Sciences, University of Oxford, Oxford, United Kingdom.
  • Friend PJ; Institute of Biomedical Engineering, University of Oxford, Oxford, United Kingdom.
  • Rodgers CT; Nuffield Department of Surgical Sciences, University of Oxford, Oxford, United Kingdom.
Magn Reson Med ; 86(6): 3246-3258, 2021 12.
Article em En | MEDLINE | ID: mdl-34272767
ABSTRACT

PURPOSE:

A shortage of suitable donor livers is driving increased use of higher risk livers for transplantation. However, current biomarkers are not sensitive and specific enough to predict posttransplant liver function. This is limiting the expansion of the donor pool. Therefore, better noninvasive tests are required to determine which livers will function following implantation and hence can be safely transplanted. This study assesses the temperature sensitivity of proton density fat fraction and relaxometry parameters and examines their potential for assessment of liver function ex vivo.

METHODS:

Six ex vivo human livers were scanned during static cold storage following normothermic machine perfusion. Proton density fat fraction, T1 , T2 , and T2∗ were measured repeatedly during cooling on ice. Temperature corrections were derived from these measurements for the parameters that showed significant variation with temperature.

RESULTS:

Strong linear temperature sensitivities were observed for proton density fat fraction (R2 = 0.61, P < .001) and T1 (R2 = 0.78, P < .001). Temperature correction according to a linear model reduced the coefficient of repeatability in these measurements by 41% and 36%, respectively. No temperature dependence was observed in T2 or T2∗ measurements. Comparing livers deemed functional and nonfunctional during normothermic machine perfusion by hemodynamic and biochemical criteria, T1 differed significantly 516 ± 50 ms for functional versus 679 ± 60 ms for nonfunctional, P = .02.

CONCLUSION:

Temperature correction is essential for robust measurement of proton density fat fraction and T1 in cold-stored human livers. These parameters may provide a noninvasive measure of viability for transplantation.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Transplante de Fígado / Fígado Gorduroso Idioma: En Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Reino Unido

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Transplante de Fígado / Fígado Gorduroso Idioma: En Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Reino Unido