Emphysema Quantification Using Ultra-Low-Dose Chest CT: Efficacy of Deep Learning-Based Image Reconstruction.
Yeom, Jeong-A; Kim, Ki-Uk; Hwang, Minhee; Lee, Ji-Won; Kim, Kun-Il; Song, You-Seon; Lee, In-Sook; Jeong, Yeon-Joo.
Afiliação
  • Yeom JA; Department of Radiology, Pusan National University Yangsan Hospital, Pusan National University School of Medicine, Yangsan 50612, Korea.
  • Kim KU; Department of Internal Medicine and Biomedical Research Institute, Pusan National University Hospital, Busan 49241, Korea.
  • Hwang M; Department of Radiology and Biomedical Research Institute, Pusan National University Hospital, Pusan National University School of Medicine, Busan 49241, Korea.
  • Lee JW; Department of Radiology and Biomedical Research Institute, Pusan National University Hospital, Pusan National University School of Medicine, Busan 49241, Korea.
  • Kim KI; Department of Radiology, Pusan National University Yangsan Hospital, Pusan National University School of Medicine, Yangsan 50612, Korea.
  • Song YS; Department of Radiology and Biomedical Research Institute, Pusan National University Hospital, Pusan National University School of Medicine, Busan 49241, Korea.
  • Lee IS; Department of Radiology and Biomedical Research Institute, Pusan National University Hospital, Pusan National University School of Medicine, Busan 49241, Korea.
  • Jeong YJ; Department of Radiology and Biomedical Research Institute, Pusan National University Hospital, Pusan National University School of Medicine, Busan 49241, Korea.
Medicina (Kaunas) ; 58(7)2022 Jul 15.
Article em En | MEDLINE | ID: mdl-35888658
ABSTRACT
Background and

Objectives:

Although reducing the radiation dose level is important during diagnostic computed tomography (CT) applications, effective image quality enhancement strategies are crucial to compensate for the degradation that is caused by a dose reduction. We performed this prospective study to quantify emphysema on ultra-low-dose CT images that were reconstructed using deep learning-based image reconstruction (DLIR) algorithms, and compared and evaluated the accuracies of DLIR algorithms versus standard-dose CT. Materials and

Methods:

A total of 32 patients were prospectively enrolled, and all underwent standard-dose and ultra-low-dose (120 kVp; CTDIvol < 0.7 mGy) chest CT scans at the same time in a single examination. A total of six image datasets (filtered back projection (FBP) for standard-dose CT, and FBP, adaptive statistical iterative reconstruction (ASIR-V) 50%, DLIR-low, DLIR-medium, DLIR-high for ultra-low-dose CT) were reconstructed for each patient. Image noise values, emphysema indices, total lung volumes, and mean lung attenuations were measured in the six image datasets and compared (one-way repeated measures ANOVA).

Results:

The mean effective doses for standard-dose and ultra-low-dose CT scans were 3.43 ± 0.57 mSv and 0.39 ± 0.03 mSv, respectively (p < 0.001). The total lung volume and mean lung attenuation of five image datasets of ultra-low-dose CT scans, emphysema indices of ultra-low-dose CT scans reconstructed using ASIR-V 50 or DLIR-low, and the image noise of ultra-low-dose CT scans that were reconstructed using DLIR-low were not different from those of standard-dose CT scans.

Conclusions:

Ultra-low-dose CT images that were reconstructed using DLIR-low were found to be useful for emphysema quantification at a radiation dose of only 11% of that required for standard-dose CT.