Your browser doesn't support javascript.
loading
Performance evaluation of a computer-aided polyp detection system with artificial intelligence for colonoscopy.
Chino, Akiko; Ide, Daisuke; Abe, Seiichiro; Yoshinaga, Shigetaka; Ichimasa, Katsuro; Kudo, Toyoki; Ninomiya, Yuki; Oka, Shiro; Tanaka, Shinji; Igarashi, Masahiro.
Afiliação
  • Chino A; Department of Gastroenterology, Cancer Institute Hospital of Japanese Foundation for Cancer Research, Tokyo, Japan.
  • Ide D; Department of Gastroenterology, Cancer Institute Hospital of Japanese Foundation for Cancer Research, Tokyo, Japan.
  • Abe S; Endoscopy Division, National Cancer Center Hospital, Tokyo, Japan.
  • Yoshinaga S; Endoscopy Division, National Cancer Center Hospital, Tokyo, Japan.
  • Ichimasa K; Digestive Disease Center, Showa University Northern Yokohama Hospital, Kanagawa, Japan.
  • Kudo T; Tokyo Endoscopic Clinic, Tokyo, Japan.
  • Ninomiya Y; Digestive Disease Center, Showa University Northern Yokohama Hospital, Kanagawa, Japan.
  • Oka S; Department of Endoscopy, Hiroshima University Hospital, Hiroshima, Japan.
  • Tanaka S; Department of Endoscopy, Hiroshima University Hospital, Hiroshima, Japan.
  • Igarashi M; Department of Endoscopy, Hiroshima University Hospital, Hiroshima, Japan.
Dig Endosc ; 36(2): 185-194, 2024 Feb.
Article em En | MEDLINE | ID: mdl-37099623
ABSTRACT

OBJECTIVES:

A computer-aided detection (CAD) system was developed to support the detection of colorectal lesions by deep learning using video images of lesions and normal mucosa recorded during colonoscopy. The study's purpose was to evaluate the stand-alone performance of this device under blinded conditions.

METHODS:

This multicenter prospective observational study was conducted at four Japanese institutions. We used 326 videos of colonoscopies recorded with patient consent at institutions in which the Ethics Committees approved the study. The sensitivity of successful detection of the CAD system was calculated using the target lesions, which were detected by adjudicators from two facilities for each lesion appearance frame; inconsistencies were settled by consensus. Successful detection was defined as display of the detection flag on the lesion for more than 0.5 s within 3 s of appearance.

RESULTS:

Of the 556 target lesions from 185 cases, detection success sensitivity was 97.5% (95% confidence interval [CI] 95.8-98.5%). The "successful detection sensitivity per colonoscopy" was 93% (95% CI 88.3-95.8%). For the frame-based sensitivity, specificity, positive predictive value, and negative predictive value were 86.6% (95% CI 84.8-88.4%), 84.7% (95% CI 83.8-85.6%), 34.9% (95% CI 32.3-37.4%), and 98.2% (95% CI 97.8-98.5%), respectively. TRIAL REGISTRATION University Hospital Medical Information Network (UMIN000044622).
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Neoplasias Colorretais / Pólipos do Colo Idioma: En Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Japão

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Neoplasias Colorretais / Pólipos do Colo Idioma: En Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Japão