B1 inhomogeneity-corrected T1 mapping and quantitative magnetization transfer imaging via simultaneously estimating Bloch-Siegert shift and magnetization transfer effects.
Jang, Albert; Han, Paul K; Ma, Chao; El Fakhri, Georges; Wang, Nian; Samsonov, Alexey; Liu, Fang.
Afiliação
  • Jang A; Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, Massachusetts, USA.
  • Han PK; Harvard Medical School, Boston, Massachusetts, USA.
  • Ma C; Harvard Medical School, Boston, Massachusetts, USA.
  • El Fakhri G; Gordon Center for Medical Imaging, Massachusetts General Hospital, Boston, Massachusetts, USA.
  • Wang N; Harvard Medical School, Boston, Massachusetts, USA.
  • Samsonov A; Gordon Center for Medical Imaging, Massachusetts General Hospital, Boston, Massachusetts, USA.
  • Liu F; Harvard Medical School, Boston, Massachusetts, USA.
Magn Reson Med ; 90(5): 1859-1873, 2023 11.
Article em En | MEDLINE | ID: mdl-37427533
PURPOSE: To introduce a method of inducing Bloch-Siegert shift and magnetization Transfer Simultaneously (BTS) and demonstrate its utilization for measuring binary spin-bath model parameters free pool spin-lattice relaxation ( T 1 F $$ {T}_1^{\mathrm{F}} $$ ), macromolecular fraction ( f $$ f $$ ), magnetization exchange rate ( k F $$ {k}_{\mathrm{F}} $$ ) and local transmit field ( B 1 + $$ {B}_1^{+} $$ ). THEORY AND METHODS: Bloch-Siegert shift and magnetization transfer is simultaneously induced through the application of off-resonance irradiation in between excitation and acquisition of an RF-spoiled gradient-echo scheme. Applying the binary spin-bath model, an analytical signal equation is derived and verified through Bloch simulations. Monte Carlo simulations were performed to analyze the method's performance. The estimation of the binary spin-bath parameters with B 1 + $$ {B}_1^{+} $$ compensation was further investigated through experiments, both ex vivo and in vivo. RESULTS: Comparing BTS with existing methods, simulations showed that existing methods can significantly bias T 1 $$ {T}_1 $$ estimation when not accounting for transmit B 1 $$ {B}_1 $$ heterogeneity and MT effects that are present. Phantom experiments further showed that the degree of this bias increases with increasing macromolecular proton fraction. Multi-parameter fit results from an in vivo brain study generated values in agreement with previous literature. Based on these studies, we confirmed that BTS is a robust method for estimating the binary spin-bath parameters in macromolecule-rich environments, even in the presence of B 1 + $$ {B}_1^{+} $$ inhomogeneity. CONCLUSION: A method of estimating Bloch-Siegert shift and magnetization transfer effect has been developed and validated. Both simulations and experiments confirmed that BTS can estimate spin-bath parameters ( T 1 F $$ {T}_1^{\mathrm{F}} $$ , f $$ f $$ , k F $$ {k}_{\mathrm{F}} $$ ) that are free from B 1 + $$ {B}_1^{+} $$ bias.