Iodine contrast volume reduction in preoperative transcatheter aortic valve implantation computed tomography: Comparison with 64- and 256-multidetector row computed tomography.
Yoshiura, T; Masuda, T; Kobayashi, Y; Kikuhara, Y; Ishibashi, T; Nonaka, H; Oku, T; Sato, T; Funama, Y.
Afiliação
  • Yoshiura T; Graduate School of Health Sciences, Kumamoto University, Kuhonji 4-24-1, Chuo-ku, Kumamoto, 860-0976, Japan; Department of Medical Technology, Tsuchiya General Hospital, Nakajima-cho 3-30, Naka-ku, Hiroshima, 730-8655, Japan. Electronic address: momotaro137koko@yahoo.co.jp.
  • Masuda T; Department of Radiological Technology, Faculty of Health Science and Technology, Kawasaki University of Medical Welfare, Matsushima 288, Okayama, 701-0193, Japan. Electronic address: takanorimasuda@yahoo.co.jp.
  • Kobayashi Y; Department of Medical Technology, Tsuchiya General Hospital, Nakajima-cho 3-30, Naka-ku, Hiroshima, 730-8655, Japan. Electronic address: kobalislin5@kih.biglobe.ne.jp.
  • Kikuhara Y; Department of Medical Technology, Tsuchiya General Hospital, Nakajima-cho 3-30, Naka-ku, Hiroshima, 730-8655, Japan. Electronic address: yamayuka0321@gmail.com.
  • Ishibashi T; Department of Medical Technology, Tsuchiya General Hospital, Nakajima-cho 3-30, Naka-ku, Hiroshima, 730-8655, Japan. Electronic address: i_toru_1221@yahoo.co.jp.
  • Nonaka H; Department of Medical Technology, Tsuchiya General Hospital, Nakajima-cho 3-30, Naka-ku, Hiroshima, 730-8655, Japan. Electronic address: swing_by_byp@yahoo.co.jp.
  • Oku T; Department of Medical Technology, Tsuchiya General Hospital, Nakajima-cho 3-30, Naka-ku, Hiroshima, 730-8655, Japan. Electronic address: pixy241724@yahoo.co.jp.
  • Sato T; Department of Medical Technology, Tsuchiya General Hospital, Nakajima-cho 3-30, Naka-ku, Hiroshima, 730-8655, Japan. Electronic address: tomoyasu_satou@yahoo.co.jp.
  • Funama Y; Department of Medical Radiation Sciences, Faculty of Life Sciences, Kumamoto University, Honjo 1-1-1, Chuo-ku, Kumamoto, 860-8556, Japan. Electronic address: funama@kumamoto-u.ac.jp.
Radiography (Lond) ; 30(1): 408-415, 2024 01.
Article em En | MEDLINE | ID: mdl-38176131
ABSTRACT

INTRODUCTION:

This study aimed to compare the vascular enhancement and radiation dose in preoperative transcatheter aortic valve implantation (TAVI) computed tomography (CT) with a reduced contrast medium (CM) using volume scans in 256-multidetector row CT (MDCT) with a standard CM using 64-MDCT.

METHODS:

This study included 78 patients with preoperative TAVI CT with either 64- or 256-MDCT. The CM was injected at 1.5 mL/kg in the 64-MDCT group and 1.0 mL/kg in the 256-MDCT group. We compared vascular enhancement of the aortic root and access routes, image quality (IQ) scores, and radiation dose in both groups.

RESULTS:

Despite the reduced CM (by 33 %) in the 256-MDCT group, the mean vascular enhancement of the right and left subclavian arteries was significantly higher than that in the 64-MDCT group [284 and 267 Hounsfield units (HU) vs. 376 and 359 HU; p < 0.05]; however, no significant differences in the mean vascular enhancement in the ascending aorta, abdominal aorta at the celiac level, and bilateral common femoral arteries were observed between the two groups (p > 0.05 for all). The median IQ scores at the aortic root were higher in the 256-MDCT group than in the 64-MDCT group (3 vs. 4; p < 0.05), and those at the femoral access routes were comparable (4 vs. 4; p = 0.33). The mean effective dose was significantly reduced by 30 % in the 256-MDCT group (23.6 vs. 16.3 mSv; p < 0.05).

CONCLUSION:

In preoperative TAVI CT, volume scans using 256-MDCT provide comparable or better vascular enhancement and IQ with a 30 % reduction in CM and radiation dose than those using 64-MDCT. IMPLICATIONS FOR PRACTICE Volume scan using 256-MDCT for preoperative TAVI CT may reduce CM and radiation dose in TAVI patients with renal dysfunction.