Performance of far forward iceless blood storage containers in controlled cold environments.
Transfusion
; 64 Suppl 2: S50-S57, 2024 May.
Article
em En
| MEDLINE
| ID: mdl-38362814
ABSTRACT
BACKGROUND:
The Golden Hour Box (GHB), an iceless blood container designed for transfusion closest to the point of injury, is used by military medical teams in remote damage control resuscitation. While its performance is well-established in hot environments, it remains underexplored in cold conditions, a significant consideration in emerging global conflict zones. STUDY DESIGN ANDMETHODS:
Four GHBs were preconditioned at +4°C or +18°C for 8 h and subsequently exposed to controlled laboratory simulated temperatures of -5, -15, and -25°C for 100 h. The study focused on their capability to maintain an internal temperature between +2 and +6°C, the recommended range for red blood cells unit storage and transport, using calibrated sensors for precise monitoring.RESULTS:
When exposed to negative Celsius temperatures, GHBs showed varied performance depending on preconditioning temperatures. When preconditioned at +4°C, GHBs maintained an internal temperature within the target range (+2 to +6°C) for 100 h at -5°C, 52 ± 1 h at -15°C, and 29 ± 4 h at -25°C. In contrast, the internal temperature of GHBs preconditioned at +18°C exceeded this range in less than 30 min, then dropped below 2°C more rapidly than those preconditioned at +4°C, occurring within 20 ± 2 h at -15 and 13 ± 1 h at -25°C.CONCLUSION:
The GHB, when properly preconditioned, effectively maintains internal temperatures suitable for blood product transport in extreme cold. Future research, including analyses of blood performances, is still needed to validate these results in more realistic operational conditions for use in cold environments.Palavras-chave
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Preservação de Sangue
/
Temperatura Baixa
Idioma:
En
Ano de publicação:
2024
Tipo de documento:
Article
País de afiliação:
França