Data Fusion of Acoustic and Optical Emission from Laser-Induced Plasma for In Situ Measurement of Rare Earth Elements in Molten LiCl-KCl.
Lee, Yunu; Foster, Richard I; Kim, Hyeongbin; Garrett, Londrea; Morgan, Bryan W; Burger, Milos; Jovanovic, Igor; Choi, Sungyeol.
Afiliação
  • Lee Y; Device Solutions, Samsung Electronics, 114 Samsung-ro, Pyeongtaek-si 17786, Republic of Korea.
  • Foster RI; Department of Nuclear and Quantum Engineering, Korea Advanced Institute of Science and Technology, 291 Daehak-ro, Yuseong-gu, Daejeon 34141, Republic of Korea.
  • Kim H; Nuclear Research Institute for Future Technology and Policy, Seoul National University, 1 Gwanak-ro, Gwanak-gu, Seoul 08826, Republic of Korea.
  • Garrett L; Department of Nuclear Engineering, Seoul National University, 1 Gwanak-ro, Gwanak-gu, Seoul 08826, Republic of Korea.
  • Morgan BW; Department of Nuclear Engineering and Radiological Sciences, University of Michigan, Ann Arbor, Michigan 48109, United States.
  • Burger M; Gérard Mourou Center for Ultrafast Optical Science, University of Michigan, Ann Arbor, Michigan 48109, United States.
  • Jovanovic I; Department of Nuclear Engineering and Radiological Sciences, University of Michigan, Ann Arbor, Michigan 48109, United States.
  • Choi S; Gérard Mourou Center for Ultrafast Optical Science, University of Michigan, Ann Arbor, Michigan 48109, United States.
Anal Chem ; 96(28): 11255-11262, 2024 Jul 16.
Article em En | MEDLINE | ID: mdl-38967238
ABSTRACT
Molten salts have a significant potential for use as next-generation nuclear reactor coolants and in pyroprocessing for the recycling of used nuclear fuel. However, the molten salt composition needs to be known at all times, and high temperatures and intense ionizing radiation pose challenges for the monitoring instrumentation. Although the technique of laser-induced breakdown spectroscopy (LIBS) has been studied for in situ measurements of molten salts, trials to improve its monitoring accuracy using chemometrics are lacking. In this study, a data fusion technique using the LIBS optical and laser-induced acoustic (LIA) signals was investigated to enhance the measurement accuracy for molten salt monitoring. Prediction models were constructed using the partial least-squares method, and the variable importance in projection scores was analyzed to evaluate the effect of incorporating the LIA signal into the analysis. This study investigates rare earth elements Eu, Er, and Pr found not only in nuclear but also in other settings such as laser and magnetic materials. The analysis of LIBS data without data fusion resulted in a root-mean-square error of prediction (RMSEP) of 0.0774-0.0913 wt %, whereas the prediction model using data fusion led to approximately 18-40% enhanced RMSEP (0.0461-0.0679 wt %). The results suggest that fusing the LIBS data with the simultaneously recorded LIA data can improve the monitoring accuracy of rare earth element composition in molten salts.