1.
J Asthma ; 43(10): 731-4, 2006 Dec.
Article in English | MEDLINE | ID: mdl-17169823

ABSTRACT

We tested reproducibility of exhaled nitric oxide (FE(NO)) and inter-operator handling when measured with a handheld device, NIOX MINO. We enrolled 20 volunteers using a priori goals of acceptable reproducibility to be mean within-subject standard deviation less than 3 parts per billion (ppb) for FE(NO) measurements less than 30 ppb, and mean coefficient of variation less than 10% for FE(NO) measurements more than 30 ppb. Seventeen subjects with measurements less than 30 ppb displayed a mean standard deviation of 1.15, and 3 subjects with FE(NO) more than 30 ppb had a mean coefficient of variation of 2.4%. We conclude that NIOX MINO demonstrates excellent reproducibility for all ranges of FE(NO).


Subject(s)
Breath Tests/instrumentation , Exhalation , Nitric Oxide , Adolescent , Adult , Child , Equipment Design , Female , Humans , Male , Middle Aged , Prospective Studies , Reproducibility of Results