In emergency cases, rapid extracorporeal membrane oxygenation (ECMO) device initialization is able to drastically reduce the incidence of patient morbidity and/or mortality. Pre-assembled and ready-to-use ECMO circuits might save up to 30-60 critical minutes in patient management. Six ECMO circuits (Oxygenator D905 EOS with REVOLUTION™ pump and Sorin PTS) were assembled in the operating room in standard conditions and then placed at 37°C for 35 days in order to evaluate possible contamination and ingrowth of micro-organisms. Every 7 days after ECMO circuit assembly and wet-priming, samples of priming fluid were analyzed to verify the presence/absence of possible common contaminants (Enterobacteriaceae, Staphylococcus aureus and fungi). Moreover, two supplementary circuits, used as positive controls, were deliberately inoculated with a known concentration of a Escherichia coli strain and prime samplings carried out at different time-points to determine bacterial growth rate. Sterility was maintained in the ECMO circuits for up to 35 days.


Subject(s)
Extracorporeal Membrane Oxygenation , Membranes, Artificial , Extracorporeal Membrane Oxygenation/instrumentation , Extracorporeal Membrane Oxygenation/methods , Time Factors