Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters

Database
Language
Affiliation country
Publication year range
1.
J Orthop Res ; 18(3): 494-9, 2000 May.
Article in English | MEDLINE | ID: mdl-10937639

ABSTRACT

We recently evaluated the peak pullout loads for anchors made from our new copolymeric swelling-type material compared with anchors made of a nonswelling material. In vitro and in vivo peak pullout loads of these anchors were evaluated after different intervals of implantation in the lateral femoral condyles of New Zealand White rabbits. Scanning electron microscopy and energy dispersive x-ray analyses were additionally performed on selected retrieved samples after pullout to examine the characteristics of bone attachment to the implant. The mean peak pullout load was greater for the swelling anchors than for the nonswelling anchors after 48 hours in vitro (46.0 +/- 15.8 compared with 10.8 +/- 9.1 N, p = 0.0541). After 2 weeks in vivo, it was significantly greater for the swelling anchors than for the nonswelling controls (177.7 +/- 41.3 compared with 53.7 +/- 17.5 N, p = 0.0024). The peak pullout load was also greater for the swelling anchors after 8 weeks in vivo; however, this difference was less pronounced than at 2 weeks (101.8 +/- 35.0 compared with 58.9 +/- 9.7 N, p = 0.0508). Furthermore, the swelling implants tended to induce bone deposition at the bone-implant interface. Results from this investigation reveal that the new family of dynamic implants has potential for applications requiring fixation to cancellous or osteoporotic bone.


Subject(s)
Bone and Bones/surgery , Prostheses and Implants , Animals , Biomechanical Phenomena , Bone and Bones/ultrastructure , Male , Microscopy, Electron, Scanning , Rabbits
SELECTION OF CITATIONS
SEARCH DETAIL