J Hosp Infect ; 17(1): 53-9, 1991 Jan.
Article in English | MEDLINE | ID: mdl-1672324

ABSTRACT

A closed sterile prefilled humidifier ('Aquapak 310') and a multiple-use humidifier ('Nebal 2') were evaluated in hospital departments to determine their susceptibility to bacterial contamination and cost. No bacterial contamination was found in the 389 samples of 'Aquapak 310' water. However, 54/164 (32.9%) samples of 'Nebal 2' water were found to be contaminated. Pseudomonas aeruginosa was the bacterium most often isolated. The cost analysis was highly influenced by the average use time. In the haemodialysis and respiratory medicine departments the average use times for the 'Aquapak 310' +/- SD were 61.6 +/- 36.2 days and 4.1 +/- 1.7 days, respectively. Using the 'Aquapak 310' system, there was a 51% financial saving in the haemodialysis department but a 2% loss in the respiratory medicine department. In these two departments we found a similar cost saving as far as staff time was concerned (88% vs. 89%). The major difference came from the cost of consumables: 26% saving in the haemodialysis department vs. 70% loss in the respiratory medicine department. Use of the prefilled sterile humidifiers represents a three-fold benefit, a lower infection risk for the patient, an important financial saving in the haemodialysis department and a decreased staff work load.


Subject(s)
Equipment Contamination , Equipment and Supplies, Hospital , Ventilators, Mechanical , Cost-Benefit Analysis , Equipment Contamination/economics , Evaluation Studies as Topic , France , Humans , Humidity , Pseudomonas aeruginosa/isolation & purification , Ventilators, Mechanical/standards
SELECTION OF CITATIONS
SEARCH DETAIL