Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Mais filtros

Base de dados
Ano de publicação
Tipo de documento
País de afiliação
Intervalo de ano de publicação
1.
2.
Nat Commun ; 11(1): 2563, 2020 05 22.
Artigo em Inglês | MEDLINE | ID: mdl-32444637

RESUMO

The increasing demand for cryo-electron microscopy (cryo-EM) reveals drawbacks in current sample preparation protocols, such as sample waste and lack of reproducibility. Here, we present several technical developments that provide efficient sample preparation for cryo-EM studies. Pin printing substantially reduces sample waste by depositing only a sub-nanoliter volume of sample on the carrier surface. Sample evaporation is mitigated by dewpoint control feedback loops. The deposited sample is vitrified by jets of cryogen followed by submersion into a cryogen bath. Because the cryogen jets cool the sample from the center, premounted autogrids can be used and loaded directly into automated cryo-EMs. We integrated these steps into a single device, named VitroJet. The device's performance was validated by resolving four standard proteins (apoferritin, GroEL, worm hemoglobin, beta-galactosidase) to ~3 Å resolution using a 200-kV electron microscope. The VitroJet offers a promising solution for improved automated sample preparation in cryo-EM studies.


Assuntos
Impressão Tridimensional , Proteínas/ultraestrutura , Manejo de Espécimes/métodos , Microscopia Crioeletrônica , Impressão Tridimensional/instrumentação , Proteínas/química , Reprodutibilidade dos Testes , Imagem Individual de Molécula , Manejo de Espécimes/instrumentação
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA