Intraluminal ultrasound applicator compatible with magnetic resonance imaging "real-time" temperature mapping for the treatment of oesophageal tumours: an ex vivo study.
Med Phys
; 31(2): 236-44, 2004 Feb.
Article
in En
| MEDLINE
| ID: mdl-15000609
High intensity ultrasound has shown considerable ability to produce precise and deep thermal coagulation necrosis. Focused, cylindrical, spherical or plane transducers have been used to induce high temperatures in tissues to coagulate proteins and kill cells. Recently magnetic resonance imaging (MRI) has been used, with extracorporeal or intracavitary focused transducers and cylindrical interstitial applicators, to monitor temperature distribution and provide feedback during heating procedures. If intraluminal applicators are used, the active part is in contact with the region of interest and it is essential to provide an accurate view of heat deposition and the extent of coagulation necrosis close to the transducer. The purpose of this study was to develop a 10 mm diameter intraluminal ultrasound applicator, designed to treat oesophageal cancers and compatible with MRI "real-time" temperature mapping. The active part of the ultrasound applicator, covered by a latex balloon, is a 15 X 8 mm2 plane transducer, which is in contact with the tumours during treatment. Each ultrasound exposure generates coagulation necrosis, in an area with the approximate shape of a rectangular parallelepiped up to 10 mm deep. When the exposures were repeated by rotating the applicator on its axis, sector-based or cylindrical volumes of necrosis could be produced, matching the shape of oesophageal cancers. Ex vivo trials were performed to demonstrate the applicator's compatibility with a clinical MRI scanner (1.5 T). MRI signals were acquired without any magnetic susceptibility distortion, even close to the applicator. Fast (0.72 images per second) 2D temperature mapping was performed during ultrasound exposure, using temperature-related proton resonance frequency shift at a resolution of 0.5 degrees C. Coagulation necrosis viewed with inversion recovery sequences, were in good agreement with the qualitative macroscopic observations made for the few cases tested in this study.
Search on Google
Collection:
01-internacional
Database:
MEDLINE
Main subject:
Ultrasonic Therapy
/
Esophageal Neoplasms
/
Magnetic Resonance Imaging
Type of study:
Qualitative_research
Limits:
Animals
Language:
En
Journal:
Med Phys
Year:
2004
Type:
Article
Affiliation country:
France