Low cost and batteryless sensor-enabled radio frequency identification tag based approaches to identify patient bed entry and exit posture transitions.
Ranasinghe, D C; Shinmoto Torres, R L; Hill, K; Visvanathan, R.
Affiliation
  • Ranasinghe DC; School of Computer Science, University of Adelaide, Australia; Auto-ID Lab, Faculty of Engineering Computer and Mathematical Sciences, University of Adelaide, SA 5000, Australia. Electronic address: damith.ranasinghe@adelaide.edu.au.
Gait Posture ; 39(1): 118-23, 2014 Jan.
Article in En | MEDLINE | ID: mdl-23850327
ABSTRACT

INTRODUCTION:

Falls in hospitals and residential care facilities commonly occur near the bed. The aim of this study was to investigate the accuracy of a continuously wearable, batteryless, low power and low cost monitoring device (Wearable Wireless Identification and Sensing Platform) with a single kinematic sensor capable of real-time monitoring to automatically detect bed entry and exit events. MATERIALS AND

METHODS:

Three dimensional acceleration readings and the strength of the transmitted signal from the WISP was interpreted to identify bed exit events and sensitivity, specificity and Receiving Operator Curves (ROC) were determined.

RESULTS:

The sensor located over sternum method performed best with sensitivity and specificity values of 92.8% and 97.5% respectively for detecting bed entry and values of 90.4% and 93.80% respectively for bed exit. On the other hand, the sensor-on-mattress algorithm achieved sensitivity and specificity values of 84.2% and 97.4% respectively for bed entry and 79% and 97.4% for bed exit detection.

CONCLUSION:

The WISP located over the sternum method is the preferred method to detect bed entry and exit. However, further work in frail older people is required to confirm the performance of this method.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Accidental Falls / Algorithms / Postural Balance / Radio Frequency Identification Device / Acceleration Type of study: Diagnostic_studies / Health_economic_evaluation / Prognostic_studies Limits: Aged / Humans Language: En Journal: Gait Posture Journal subject: ORTOPEDIA Year: 2014 Type: Article

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Accidental Falls / Algorithms / Postural Balance / Radio Frequency Identification Device / Acceleration Type of study: Diagnostic_studies / Health_economic_evaluation / Prognostic_studies Limits: Aged / Humans Language: En Journal: Gait Posture Journal subject: ORTOPEDIA Year: 2014 Type: Article