Your browser doesn't support javascript.
loading
Evaluation of an innovative pediatric isolation (PI) bed using fluid dynamics simulation and aerosol isolation efficacy.
Liu, Tiantian; Guo, Yubing; Hao, Xiaotang; Wang, Mei; He, Shicong; Lin, Zhengshi; Zhou, Rong.
Affiliation
  • Liu T; School of Public Health, Guangdong Pharmaceutical University, Guangzhou, Guangdong, 510310 China.
  • Guo Y; State Key Laboratory of Respiratory Disease, National Clinical Research Center for Respiratory Disease, Guangzhou Institute of Respiratory Health, First Affiliated Hospital of Guangzhou Medical University, Guangzhou Medical University, Guangzhou, China.
  • Hao X; School of Public Health, Guangdong Pharmaceutical University, Guangzhou, Guangdong, 510310 China.
  • Wang M; School of Public Health, Guangdong Pharmaceutical University, Guangzhou, Guangdong, 510310 China.
  • He S; State Key Laboratory of Respiratory Disease, National Clinical Research Center for Respiratory Disease, Guangzhou Institute of Respiratory Health, First Affiliated Hospital of Guangzhou Medical University, Guangzhou Medical University, Guangzhou, China.
  • Lin Z; State Key Laboratory of Respiratory Disease, National Clinical Research Center for Respiratory Disease, Guangzhou Institute of Respiratory Health, First Affiliated Hospital of Guangzhou Medical University, Guangzhou Medical University, Guangzhou, China.
  • Zhou R; State Key Laboratory of Respiratory Disease, National Clinical Research Center for Respiratory Disease, Guangzhou Institute of Respiratory Health, First Affiliated Hospital of Guangzhou Medical University, Guangzhou Medical University, Guangzhou, China.
Build Simul ; 14(5): 1543-1552, 2021.
Article in En | MEDLINE | ID: mdl-33686353
ABSTRACT
Airborne transmission is an important mechanism of spread for both viruses and bacteria in hospitals, with nosocomial infections putting a great burden on public health. In this study, we designed and manufactured a bed for pediatric clinic consultation rooms providing air isolation to protect patients and medical personnel from pathogen transmission. The pediatric isolation bed has several primary efficiency filters and a high-efficiency particulate air filter in the bedside unit. The air circulation between inlet and outlet forms negative pressure to remove the patient's exhaled air timeously and effectively. A computational fluid dynamics model was used to calculate the speed of the airflow and the angle of sampler. Following this, we conducted purification experiments using cigarette smoke, Staphylococcus albus (S. albus) and human adenovirus type 5 (HAdV-5) to demonstrate the isolation efficacy. The results showed that the patient's head should be placed as close to the air inlet hood as possible, and an air intake wind speed of 0.86 m/s was effective. The isolation efficacy of the pediatric isolation bed was demonstrated by computational fluid dynamics technology. The isolation efficiency against cigarette smoke exceeded 91.8%, and against S. albus was greater than 99.8%, while the isolation efficiency against HAdV-5 was 100%. The pediatric isolation bed could be used where isolation wards are unavailable, such as in intensive care units and primary clinical settings, to control hospital acquired infections.
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Build Simul Year: 2021 Type: Article