Comparison of four clinically validated testosterone LC-MS/MS assays: Harmonization is an attainable goal.
French, Deborah; Drees, Julia; Stone, Judith A; Holmes, Daniel T; van der Gugten, J Grace.
Affiliation
  • French D; Department of Laboratory Medicine, University of California San Francisco, San Francisco, CA 94107, United States.
  • Drees J; Kaiser Permanente Northern California Regional Laboratory, Richmond, CA, United States.
  • Stone JA; University of California San Diego Health Center for Advanced Laboratory Medicine, San Diego, CA, United States.
  • Holmes DT; Department of Pathology and Laboratory Medicine, St Paul's Hospital, Vancouver, British Columbia, Canada.
  • van der Gugten JG; Department of Pathology and Laboratory Medicine, University of British Columbia, Vancouver, British Columbia, Canada.
Clin Mass Spectrom ; 11: 12-20, 2019 Jan.
Article in En | MEDLINE | ID: mdl-34841068
ABSTRACT

INTRODUCTION:

Immunoassays and liquid chromatography-tandem mass spectrometry assays are commonly employed in clinical laboratories for measurement of total testosterone in serum. Results obtained from either of these methodologies compare poorly due to differences in calibration and/or inadvertent detection of interfering substances by the immunoassays. Standardization efforts are underway, but recent studies indicate that accuracy remains an issue.

METHODS:

This study compares the results from four independently developed and validated LC-MS/MS assays for total testosterone. The calibration for each assay was verified using National Institute of Standards and Technology Standard Reference Material 971.

RESULTS:

Initially, one of the four assays had a mean percent difference of +11.44%, compared to the All Method Mean, but following re-verification of all five non-zero calibrator concentrations with the NIST SRM 971, the mean percent difference decreased to -4.88%. Subsequently, the agreement between all four assays showed a mean bias of <5% across the range of all testosterone concentrations (0.13-38.10 nmol/L; 3.7-1098 ng/dL), including at low concentrations of <1 nmol/L (<29 ng/dL).

CONCLUSIONS:

Excellent agreement between four independently developed LC-MS/MS assays demonstrates that harmonization using standard reference material is attainable. However, as we found in this study, to ensure accurate calibration it is critical to validate the concentrations of new lots of calibrators.
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Type of study: Guideline Language: En Journal: Clin Mass Spectrom Year: 2019 Type: Article Affiliation country: United States