Independent Monte Carlo dose calculation identifies single isocenter multi-target radiosurgery targets most likely to fail pre-treatment measurement.
J Appl Clin Med Phys
; 25(6): e14290, 2024 Jun.
Article
in En
| MEDLINE
| ID: mdl-38289874
ABSTRACT
PURPOSE:
For individual targets of single isocenter multi-target (SIMT) Stereotactic radiosurgery (SRS), we assess dose difference between the treatment planning system (TPS) and independent Monte Carlo (MC), and demonstrate persistence into the pre-treatment Quality Assurance (QA) measurement.METHODS:
Treatment plans from 31 SIMT SRS patients were recalculated in a series of scenarios designed to investigate sources of discrepancy between TPS and independent MC. Targets with > 5% discrepancy in DMean[Gy] after progressing through all scenarios were measured with SRS MapCHECK. A matched pair analysis was performed comparing SRS MapCHECK results for these targets with matched targets having similar characteristics (volume & distance from isocenter) but no such MC dose discrepancy.RESULTS:
Of 217 targets analyzed, individual target mean dose (DMean[Gy]) fell outside a 5% threshold for 28 and 24 targets before and after removing tissue heterogeneity effects, respectively, while only 5 exceeded the threshold after removing effect of patient geometry (via calculation on StereoPHAN geometry). Significant factors affecting agreement between the TPS and MC included target distance from isocenter (0.83% decrease in DMean[Gy] per 2 cm), volume (0.15% increase per cc), and degree of plan modulation (0.37% increase per 0.01 increase in modulation complexity score). SRS MapCHECK measurement had better agreement with MC than with TPS (2%/1 mm / 10% threshold gamma pass rate (GPR) = 99.4 ± 1.9% vs. 93.1 ± 13.9%, respectively). In the matched pair analysis, targets exceeding 5% for MC versus TPS also had larger discrepancies between TPS and measurement with no GPR (2%/1 mm / 10% threshold) exceeding 90% (71.5% ± 16.1%); whereas GPR was high for matched targets with no such MC versus TPS difference (96.5% ± 3.3%, p = 0.01).CONCLUSIONS:
Independent MC complements pre-treatment QA measurement for SIMT SRS by identifying problematic individual targets prior to pre-treatment measurement, thus enabling plan modifications earlier in the planning process and guiding selection of targets for pre-treatment QA measurement.Key words
Full text:
1
Collection:
01-internacional
Database:
MEDLINE
Main subject:
Quality Assurance, Health Care
/
Radiotherapy Dosage
/
Radiotherapy Planning, Computer-Assisted
/
Monte Carlo Method
/
Radiosurgery
/
Radiotherapy, Intensity-Modulated
Type of study:
Prognostic_studies
Limits:
Humans
Language:
En
Journal:
J Appl Clin Med Phys
Journal subject:
BIOFISICA
Year:
2024
Type:
Article
Affiliation country: