[Research on the method of loading exosomes onto absorbable stents].
Wang, Chenchen; Li, Tianbo; Guo, Ruimin; Ji, Qinghui; Wang, Jiangning; Cao, Yulin.
Affiliation
  • Wang C; Jiamusi University, Jiamusi Heilongjiang, 154007, P. R. China.
  • Li T; .
  • Guo R; .
  • Ji Q; Jiamusi University, Jiamusi Heilongjiang, 154007, P. R. China.
  • Wang J; Department of Orthopedics, Shijitan Hospital Affiliated to Capital Medical University, Beijing, 100038, P. R. China.
  • Cao Y; Tangyi Holdings (Shenzhen) Co., Ltd, Shenzhen Guangdong, 518000, P. R. China.
Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi ; 38(2): 206-210, 2024 Feb 15.
Article in Zh | MEDLINE | ID: mdl-38385234
ABSTRACT

Objective:

To explore a method of loading exosomes onto absorbable stents.

Methods:

By building a stent-(3-aminopropyl) triethoxysilane-1, 2-distearoyl-sn-glycero-3-phosphoethanolamine-N-[methoxy (polyethylene glycol) 5000]-exosomes connection, the exosomes were loaded onto absorbable stents to obtained the exosome-eluting absorbable stents. The surface conditions of the stents and absorption of exosomes were observed by scanning electron microscope and identified through the time-of-flight mass spectrometry; the roughness of the stents' surfaces was observed by atomic force microscope; the appearances and sizes of the stents were observed by stereomicroscope; and the radial force was tested by tensile test machine. The absorbable stents were used as control.

Results:

The scanning electron microscope observation showed that the exosome-eluting absorbable stents had some small irregular cracks on the surface where many exosomes could be seen. The atomic force microscopy observation showed that within the range of 5 µm 2, the surface roughness of the absorbable stents was ±20 nm, while the surface roughness of the exosome-eluting absorbable stents was ±70 nm. In the results of time-of-flight mass spectrometry, both the exosome-eluting absorbable stents and exosomes had a peak at the mass charge ratio of 81 (m/z 81), while the absorbable stents did not have this peak. The peak of exosome-eluting absorbable stents at m/z 73 showed a significant decrease compared to the absorbable stents. The stereomicroscope observation showed that the sizes of exosome-eluting absorbable stents met standards and the surfaces had no cracks, burrs, or depressions. The radial force results of the exosome-eluting absorbable stents met the strength standards of the original absorbable stent.

Conclusion:

By applying the chemical connection method, the exosomes successfully loaded onto the absorbable stents. And the sizes and radial forces of this exosome-eluting absorbable stents meet the standards of the original absorbable stents.