Your browser doesn't support javascript.
loading
Point of care hematocrit and hemoglobin in cardiac surgery: a review.
Myers, Gerard J; Browne, Joe.
Afiliação
  • Myers GJ; Cardiovascular Perfusion, QEII & IWK Health Science Centers, Halifax, Nova Scotia, Canada. edit@ns.sympatico.ca
Perfusion ; 22(3): 179-83, 2007 May.
Article em En | MEDLINE | ID: mdl-18018397
The use of point-of-care blood gas analyzers in cardiac surgery has been on the increase over the past decade. The availability of these analyzers in the operating room and post-operative intensive care units eliminates the time delays to transport samples to the main laboratory and reduces the amount of blood sampled to measure such parameters as electrolytes, blood gases, lactates, glucose and hemoglobin/hematocrit. Point-of-care analyzers also lead to faster and more reliable clinical decisions while the patient is still on the heart lung machine. Point-of-care devices were designed to provide safe, appropriate and consistent care of those patients in need of rapid acid/base balance and electrolyte management in the clinical setting. As a result, clinicians rely on their values to make decisions regarding ventilation, acid/base management, transfusion and glucose management. Therefore, accuracy and reliability are an absolute must for these bedside analyzers in both the cardiac operating room and the post-op intensive care units. Clinicians have a choice of two types of technology to measure hemoglobin/hematocrit during bypass, which subsequently determines their patient's level of hemodilution, as well as their transfusion threshold. All modern point-of-care blood gas analyzers measure hematocrit using a technology called conductivity, while other similar devices measure hemoglobin using a technology called co-oximetry. The two methods are analyzed and compared in this review. The literature indicates that using conductivity to measure hematocrit during and after cardiac surgery could produce inaccurate results when hematocrits are less than 30%, and, therefore, result in unnecessary homologous red cell transfusions in some patients. These inaccuracies are influenced by several factors that are common and unique to cardiopulmonary bypass, and will also be reviewed here. It appears that the only accurate, consistent and reliable method to determine hemodilution and establish transfusion thresholds based on nadir hematocrits during cardiopulmonary bypass, and immediately post cardiac surgery, is with the use of co-oximetry.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Hemoglobinas / Sistemas Automatizados de Assistência Junto ao Leito / Procedimentos Cirúrgicos Cardíacos / Hematócrito / Cuidados Intraoperatórios Tipo de estudo: Diagnostic_studies / Etiology_studies / Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: Perfusion Assunto da revista: CARDIOLOGIA Ano de publicação: 2007 Tipo de documento: Article País de afiliação: Canadá
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Hemoglobinas / Sistemas Automatizados de Assistência Junto ao Leito / Procedimentos Cirúrgicos Cardíacos / Hematócrito / Cuidados Intraoperatórios Tipo de estudo: Diagnostic_studies / Etiology_studies / Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: Perfusion Assunto da revista: CARDIOLOGIA Ano de publicação: 2007 Tipo de documento: Article País de afiliação: Canadá