Device closure of ventricular septal defects by hybrid procedures: a multicenter retrospective study.
Michel-Behnke, Ina; Ewert, Peter; Koch, Andreas; Bertram, Harald; Emmel, Mathias; Fischer, Gunther; Gitter, Roland; Kozlik-Feldman, Rainer; Motz, Reinald; Kitzmüller, Erwin; Kretschmar, Oliver.
Afiliação
  • Michel-Behnke I; Department of Pediatric Cardiology, Pediatric Heart Center, Giessen, Germany. ina.michel-behnke@meduniwien.ac.at
Catheter Cardiovasc Interv ; 77(2): 242-51, 2011 Feb 01.
Article em En | MEDLINE | ID: mdl-20517999
ABSTRACT

OBJECTIVE:

This study reports on the feasibility, efficacy, and outcome of hybrid procedures to close ventricular septal defects (VSD), reflecting the experience of 11 centers in Germany, Austria, and Switzerland.

BACKGROUND:

Beating heart closure of VSD has attracted interest in small infants, complex anomalies and postinfarction scenarios where patients are at high risk during surgery. Perventricular or intraoperative device placement allows access to the lesions where percutaneous delivery is limited.

METHODS:

Between December 2001 and April 2009, placement of Amplatzer septal occluders was attempted in 26 patients. The defects were located in the perimembranous (n = 5) and muscular septum (n = 21). In 20 patients, a perventricular approach was used, and, in six, the occluders were placed under direct visualization being part of a complex heart surgery.

RESULTS:

In 23 of 26 procedures, device placement was successful (88.5%). The mean defect size was 7.8 mm (range, 3.5-20). The occluder types were perimembranous VSD occluder (n = 4), muscular VSD occluder (n = 20), postinfarct VSD occluder (n = 1), and ASD occluder (n = 1) with a ratio device/defect of 0.9-2.4 (median 1.15). Device removal was necessary in three due to arrhythmia, malpositioning, and additional defects. Pericardial effusion occurred once. In the remaining 22 patients, there were no procedure or device-related complications. During mean follow up of 1.4 years (range, 1 day-3.9 years), a residual shunt that was more than trivial was observed in one patient out of 21 successful procedures.

CONCLUSIONS:

Perventricular and intraoperative device closure of VSD is as effective as a surgical patch and averts the increased morbidity of conventional surgical repair in a subgroup of high-risk patients.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Cateterismo Cardíaco / Dispositivo para Oclusão Septal / Comunicação Interventricular Tipo de estudo: Etiology_studies / Observational_studies / Risk_factors_studies Limite: Aged / Child, preschool / Humans / Infant / Newborn País/Região como assunto: Europa Idioma: En Revista: Catheter Cardiovasc Interv Assunto da revista: CARDIOLOGIA Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Cateterismo Cardíaco / Dispositivo para Oclusão Septal / Comunicação Interventricular Tipo de estudo: Etiology_studies / Observational_studies / Risk_factors_studies Limite: Aged / Child, preschool / Humans / Infant / Newborn País/Região como assunto: Europa Idioma: En Revista: Catheter Cardiovasc Interv Assunto da revista: CARDIOLOGIA Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Alemanha