Your browser doesn't support javascript.
loading
Comparison of two methods for blood lead analysis in cattle: graphite-furnace atomic absorption spectrometry and LeadCare(R) II system.
Bischoff, Karyn; Gaskill, Cynthia; Erb, Hollis N; Ebel, Joseph G; Hillebrandt, Joseph.
Afiliação
  • Bischoff K; New York State Animal Health Diagnostic Center, Cornell University, Population Medicines and Sciences Health and Diagnostic Services, Ithaca, NY 14853, USA. klb72@cornell.edu
J Vet Diagn Invest ; 22(5): 729-33, 2010 Sep.
Article em En | MEDLINE | ID: mdl-20807929
ABSTRACT
The current study compared the LeadCare(R) II test kit system with graphite-furnace atomic absorption spectrometry for blood lead (Pb) analysis in 56 cattle accidentally exposed to Pb in the field. Blood Pb concentrations were determined by LeadCare II within 4 hr of collection and after 72 hr of refrigeration. Blood Pb concentrations were determined by atomic absorption spectrometry, and samples that were coagulated (n = 12) were homogenized before analysis. There was strong rank correlation (R(2) = 0.96) between atomic absorption and LeadCare II (within 4 hr of collection), and a conversion formula was determined for values within the observed range (3-91 mcg/dl, although few had values >40 mcg/dl). Median and mean blood pb concentrations for atomic absorption were 7.7 and 15.9 mcg/dl, respectively; for LeadCare II, medians were 5.2 mcg/dl at 4 hr and 4.9 mcg/dl at 72 hr, and means were 12.4 and 11.7, respectively. LeadCare II results at 4 hr strongly correlated with 72 hr results (R(2) = 0.96), but results at 72 hr were lower (P < 0.01). There was no significant difference between coagulated and uncoagulated samples run by atomic absorption. Although there have been several articles that compared LeadCare with other analytical techniques, all were for the original system, not LeadCare II. The present study indicated that LeadCare II results correlated well with atomic absorption over a wide range of blood Pb concentrations and that refrigerating samples for up to 72 hr before LeadCare II analysis was acceptable for clinical purposes.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Bovinos / Chumbo Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Animals Idioma: En Revista: J Vet Diagn Invest Assunto da revista: MEDICINA VETERINARIA Ano de publicação: 2010 Tipo de documento: Article País de afiliação: Estados Unidos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Bovinos / Chumbo Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Animals Idioma: En Revista: J Vet Diagn Invest Assunto da revista: MEDICINA VETERINARIA Ano de publicação: 2010 Tipo de documento: Article País de afiliação: Estados Unidos