A comparative reliability and performance study of different stent designs in terms of mechanical properties: foreshortening, recoil, radial force, and flexibility.
Kim, Dong Bin; Choi, Hyuk; Joo, Sang Min; Kim, Han Ki; Shin, Jae Hee; Hwang, Min Ho; Choi, Jaesoon; Kim, Dong-Gon; Lee, Kwang Ho; Lim, Chun Hak; Yoo, Sun Kook; Lee, Heung-Man; Sun, Kyung.
Afiliação
  • Kim DB; Graduate Program in Biomedical Engineering, Yonsei University.
Artif Organs ; 37(4): 368-79, 2013 Apr.
Article em En | MEDLINE | ID: mdl-23461583
ABSTRACT
This study seeks to improve the mechanical performance of stents by conducting reliability performance testing and finite element method (FEM)-based simulations for coronary stents. Three commercially available stent designs and our own new design were tested to measure the factors affecting performance, specifically foreshortening, recoil, radial force, and flexibility. The stents used in the present experiments were 3 mm in working diameter and 18 mm of working length. The results of the experiments indicate that the foreshortening of stents A, B, C, and our new design, D, was equivalent to 2.25, 0.67, 0.46, and 0.41%, respectively. The recoil of stents A, B, C, and D was 6.00, 4.35, 3.50, and 4.36%, respectively. Parallel plate radial force measurements were A, 3.72 ± 0.28 N; B, 3.81 ± 0.32 N; C, 4.35 ± 0.18 N; and D, 4.02 ± 0.24 N. Radial forces determined by applying uniform pressure in the circumferential direction were A, 28.749 ± 0.81 N; B, 32.231 ± 1.80 N; C, 34.522 ± 3.06 N; and D, 42.183 ± 2.84 N. The maximum force of crimped stent at 2.2-mm deflection was 1.01 ± 0.08 N, 0.82 ± 0.08 N, 0.92 ± 0.12 N, and 0.68 ± 0.07 N for each of stents A, B, C and D. The results of this study enabled us to identify several factors to enhance the performance of stents. In comparing these stents, we found that our design, stent D, which was designed by a collaborative team from seven universities, performed better than the commercial stents across all parameter of foreshortening, recoil, radial force, and flexibility.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Stents Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Artif Organs Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Stents Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Artif Organs Ano de publicação: 2013 Tipo de documento: Article