Effect of hydroxocobalamin on surface oximetry in nonexposed humans.
Cashin, Brian V; Matlock, Aaron G; Kang, Chris; Reynolds, Penny S; Wills, Brandon K.
Afiliação
  • Cashin BV; Department of Emergency Medicine, Madigan Army Medical Center, Tacoma, WA, USA.
Prehosp Disaster Med ; 28(4): 367-9, 2013 Aug.
Article em En | MEDLINE | ID: mdl-23635811
ABSTRACT

INTRODUCTION:

The newer cyanide antidote, hydroxocobalamin, due to its pigmentation, has been found to cause interferences in some laboratory assays. Co-oximetry may also be affected by hydroxocobalamin, leading to false elevations in hemoglobin concentration, methemoglobin, carboxyhemoglobin, and false decreases in oxyhemoglobin. The Masimo Radical-7 is a medical device that performs noninvasive oximetry and estimates hemoglobin (Hb) concentration and percent carboxyhemoglobin (COHb), methemoglobin (MetHb), and oxyhemoglobin saturation (O2Hb). STUDY

OBJECTIVES:

The study sought to determine the effect of hydroxocobalamin on noninvasive measurement of hemoglobin indices using the Masimo Radical-7 monitor.

METHODS:

Seven asymptomatic volunteers who were unexposed to cyanide had baseline heart rate (HR), blood pressure (BP), and oximeter measurements recorded followed by an infusion of five grams of hydroxocobalamin over 15 minutes. The above parameters were subsequently recorded at 5, 10, 15, 30 and 60 minutes post infusion. Data were analyzed by calculating the area under the curve (AUC) for each variable and comparing the results to expected values by paired t tests. Expected AUC values were calculated by extrapolating baseline values across the entire time period.

RESULTS:

The mean differences from baseline values with 95% confidence intervals and t tests of mean difference were SBP 11 mm Hg (95% CI, 0-22; P = .051); HR -9 (95% CI, -15 to -3; P = .01); Hb -0.1 (95% CI, -0.7 to 0.4; P = .57); O2Hb 0 g/dL (95% CI, -1 to 1; P = .41); COHb -1 (95% CI, -3 to 1; P = .25); MetHb -0.2 (95% CI, -0.3 to 0; P = .03).

DISCUSSION:

After infusion of hydroxocobalamin there was a significant elevation of systolic blood pressure and decrease in heart rate. There were no significant differences in Hb, O2Hb, and COHb. Although percent methemoglobin concentrations were statistically lower, the authors feel this difference is of trivial clinical significance.

CONCLUSION:

The administration of hydroxocobalamin does not significantly impact noninvasive oximetry.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Hemoglobinas / Oximetria / Hidroxocobalamina Tipo de estudo: Observational_studies / Prognostic_studies Limite: Humans / Male País/Região como assunto: America do norte Idioma: En Revista: Prehosp Disaster Med Assunto da revista: MEDICINA DE EMERGENCIA Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Hemoglobinas / Oximetria / Hidroxocobalamina Tipo de estudo: Observational_studies / Prognostic_studies Limite: Humans / Male País/Região como assunto: America do norte Idioma: En Revista: Prehosp Disaster Med Assunto da revista: MEDICINA DE EMERGENCIA Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Estados Unidos