OBJECTIVE:
In order to improve image quality in a simultaneous fMRI-EEG study with patients suffering from the involuntary movements typical for Huntington's disease, the aim was to develop a technique for immobilizing the heads of our patients inside an MRI head coil.METHODS:
We modified a mask technique previously used for reliable repositioning in temporally fractionated radiotherapy. The mask was tested in three patients with Huntington's disease, acquiring structural and functional MR images with simultaneous EEG with and without the mask.RESULTS:
Image as well as EEG signal quality were significantly improved in patients wearing the mask. However, the image quality with mask was comparable to acquisitions from patients without movement disorders only in patients with light to moderate dyskinesia. Although image quality was also significantly improved in a patient suffering from severe dyskinesia with quasi-continuous involuntary movements, the quality of both the MR images as well as the EEG signal was lower than what would be expected in a healthy control person.CONCLUSION:
We have succeeded in developing a mask that fits into the MRI head coil, does not disturb the MRI signal, and significantly improves both fMRI and EEG signal quality.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Encéfalo
/
Imageamento por Ressonância Magnética
/
Doença de Huntington
Limite:
Humans
Idioma:
En
Revista:
Acta Neurochir Suppl
Ano de publicação:
2013
Tipo de documento:
Article