Evaluation of an automated breast 3D-ultrasound system by comparing it with hand-held ultrasound (HHUS) and mammography.
Golatta, Michael; Baggs, Christina; Schweitzer-Martin, Mirjam; Domschke, Christoph; Schott, Sarah; Harcos, Aba; Scharf, Alexander; Junkermann, Hans; Rauch, Geraldine; Rom, Joachim; Sohn, Christof; Heil, Joerg.
Afiliação
  • Golatta M; University Breast Unit, Im Neuenheimer Feld 440, 69120, Heidelberg, Germany, michael.golatta@med.uni-heidelberg.de.
Arch Gynecol Obstet ; 291(4): 889-95, 2015 Apr.
Article em En | MEDLINE | ID: mdl-25311201
ABSTRACT

PURPOSE:

Automated three-dimensional (3D) breast ultrasound (US) systems are meant to overcome the shortcomings of hand-held ultrasound (HHUS). The aim of this study is to analyze and compare clinical performance of an automated 3D-US system by comparing it with HHUS, mammography and the clinical gold standard (defined as the combination of HHUS, mammography and-if indicated-histology).

METHODS:

Nine hundred and eighty three patients (=1,966 breasts) were enrolled in this monocentric, explorative and prospective cohort study. All examinations were analyzed blinded to the patients´ history and to the results of the routine imaging. The agreement of automated 3D-US with HHUS, mammography and the gold standard was assessed with kappa statistics. Sensitivity, specificity and positive and negative predictive value were calculated to assess the test performance.

RESULTS:

Blinded to the results of the gold standard the agreement between automated 3D-US and HHUS or mammography was fair, given by a Kappa coefficient of 0.31 (95% CI [0.26;0.36], p < 0.0001) and 0.25 (95% CI [0.2;0.3], p < 0.0001), respectively. Our results showed a high negative predictive value (NPV) of 98%, a high specificity of 85% and a sensitivity of 74% based on the cases with US-guided biopsy. Including the cases where the lesion was seen in a second-look automated 3D-US the sensitivity improved to 84% (NPV = 99%, specificity = 85%).

CONCLUSION:

The results of this study let us suggest, that automated 3D-US might be a helpful new tool in breast imaging, especially in screening.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Mama / Mamografia / Ultrassonografia Mamária Tipo de estudo: Diagnostic_studies / Evaluation_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adult / Aged / Female / Humans / Middle aged Idioma: En Revista: Arch Gynecol Obstet Assunto da revista: GINECOLOGIA / OBSTETRICIA Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Mama / Mamografia / Ultrassonografia Mamária Tipo de estudo: Diagnostic_studies / Evaluation_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adult / Aged / Female / Humans / Middle aged Idioma: En Revista: Arch Gynecol Obstet Assunto da revista: GINECOLOGIA / OBSTETRICIA Ano de publicação: 2015 Tipo de documento: Article