Your browser doesn't support javascript.
loading
Effects of Detector Thickness on Geometric Sensitivity and Event Positioning Errors in the Rectangular PET/X Scanner.
MacDonald, Lawrence R; Hunter, William C J; Kinahan, Paul E; Miyaoka, Robert S.
Afiliação
  • MacDonald LR; Department of Radiology, University of Washington, Seattle, WA 98195 USA.
  • Hunter WC; Department of Radiology, University of Washington, Seattle, WA 98195 USA.
  • Kinahan PE; Department of Radiology, University of Washington, Seattle, WA 98195 USA.
  • Miyaoka RS; Department of Radiology, University of Washington, Seattle, WA 98195 USA.
IEEE Trans Nucl Sci ; 60(5): 3242-3252, 2013 Oct.
Article em En | MEDLINE | ID: mdl-26160982
We used simulations to investigate the relationship between sensitivity and spatial resolution as a function of crystal thickness in a rectangular PET scanner intended for quantitative assessment of breast cancers. The system had two 20 × 15-cm2 and two 10 × 15-cm2 flat detectors forming a box, with the larger detectors separated by 4 or 8 cm. Depth-of-interaction (DOI) resolution was modeled as a function of crystal thickness based on prior measurements. Spatial resolution was evaluated independent of image reconstruction by deriving and validating a surrogate metric from list-mode data (dFWHM). When increasing crystal thickness from 5 to 40 mm, and without using DOI information, the dFWHM for a centered point source increased from 0.72 to 1.6 mm. Including DOI information improved dFWHM by 12% and 27% for 5- and 40-mm-thick crystals, respectively. For a point source in the corner of the FOV, use of DOI information improved dFWHM by 20% (5-mm crystal) and 44% (40-mm crystal). Sensitivity was 7.7% for 10-mm-thick crystals (8-cm object). Increasing crystal thickness on the smaller side detectors from 10 to 20 mm (keeping 10-mm crystals on the larger detectors) boosted sensitivity by 24% (relative) and degraded dFWHM by only ~3%/8% with/without DOI information. The benefits of measuring DOI must be evaluated in terms of the intended clinical task of assessing tracer uptake in small lesions. Increasing crystal thickness on the smaller side detectors provides substantial sensitivity increase with minimal accompanying loss in resolution.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies Idioma: En Revista: IEEE Trans Nucl Sci Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies Idioma: En Revista: IEEE Trans Nucl Sci Ano de publicação: 2013 Tipo de documento: Article