Absolute measurement of (198)Au activity in gold foil using plastic scintillators and a well-type NaI(Tl) detector.
Appl Radiat Isot
; 109: 382-385, 2016 Mar.
Article
em En
| MEDLINE
| ID: mdl-26678344
A beta-gamma coincidence system has been developed for measuring (198)Au activity in gold foils. The system was validated by Monte Carlo simulations and by measuring the activity of a (60)Co point-source. To study effects such as self-shielding of beta particles in gold foils, (198)Au activity measurements and simulations were performed for various scintillators and foil sizes. The measured (198)Au activities were ~1% above the reference activity, which might be due to self-shielding of beta particles. The measured and simulated (198)Au activities agreed, suggesting feasibility of precise activity measurement.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Idioma:
En
Revista:
Appl Radiat Isot
Assunto da revista:
MEDICINA NUCLEAR
/
SAUDE AMBIENTAL
Ano de publicação:
2016
Tipo de documento:
Article