Your browser doesn't support javascript.
loading
Reducing the small-heart effect in pediatric gated myocardial perfusion single-photon emission computed tomography.
Yoneyama, Hiroto; Nakajima, Kenichi; Okuda, Koichi; Matsuo, Shinro; Onoguchi, Masahisa; Kinuya, Seigo; Edenbrandt, Lars.
Afiliação
  • Yoneyama H; Department of Radiological Technology, Kanazawa University Hospital, 13-1 Takara-machi, Kanazawa, 920-8641, Japan. kizu@cf6.so-net.ne.jp.
  • Nakajima K; Department of Nuclear Medicine, Kanazawa University Hospital, Kanazawa, Japan.
  • Okuda K; Department of Physics, Kanazawa Medical University, Kanazawa, Japan.
  • Matsuo S; Department of Nuclear Medicine, Kanazawa University Hospital, Kanazawa, Japan.
  • Onoguchi M; Department of Health Science, Graduate School of Medical Science, Kanazawa University, Kanazawa, Japan.
  • Kinuya S; Department of Nuclear Medicine, Kanazawa University Hospital, Kanazawa, Japan.
  • Edenbrandt L; Department of Molecular and Clinical Medicine, University of Gothenburg, Gothenburg, Sweden.
J Nucl Cardiol ; 24(4): 1378-1388, 2017 08.
Article em En | MEDLINE | ID: mdl-27197818
ABSTRACT

BACKGROUND:

We compared two reconstruction algorisms and two cardiac functional evaluation software programs in terms of their accuracy for estimating ejection fraction (EF) of small hearts (SH).

METHODS:

The study group consisted of 66 pediatric patients. Data were reconstructed using a filtered back projection (FBP) method without the resolution correction (RC) and an iterative method based on an ordered subset expectation maximization (OSEM) algorithm with the RC. EF was evaluated using two software programs of quantitative gated single-photon emission computed tomography (SPECT) (QGS) and cardioREPO. We compared the EF of gated myocardial perfusion SPECT to echocardiographic measurement (Echo).

RESULTS:

Forty-eight of 66 patients had an end-systolic volume < 20 mL which was used as the criterion for being included in the SH group, and the SH effect led to an overestimation of EF. While significant differences were observed between Echo (66.9 ± 5.0%) and QGS-FBP without RC (76.9 ± 8.4%, P < .0001), QGS-OSEM with RC (76.6 ± 8.6%, P < .0001), and cardioREPO-FBP without RC (72.1 ± 10.0%, P = .0011), no significant difference was observed between Echo and cardioREPO-OSEM with RC (67.4 ± 6.1%) in SH group.

CONCLUSIONS:

In pediatric gated myocardial perfusion SPECT, the SH effect can be significantly reduced when an OSEM algorithm is used with RC in combination with the specific cardioREPO algorithm.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Emissão de Fóton Único de Sincronização Cardíaca / Imagem de Perfusão do Miocárdio Limite: Adolescent / Child / Female / Humans / Male Idioma: En Revista: J Nucl Cardiol Assunto da revista: CARDIOLOGIA Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Japão

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Emissão de Fóton Único de Sincronização Cardíaca / Imagem de Perfusão do Miocárdio Limite: Adolescent / Child / Female / Humans / Male Idioma: En Revista: J Nucl Cardiol Assunto da revista: CARDIOLOGIA Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Japão