Your browser doesn't support javascript.
loading
Open ISEmeter: An open hardware high-impedance interface for potentiometric detection.
Salvador, C; Mesa, M S; Durán, E; Alvarez, J L; Carbajo, J; Mozo, J D.
Afiliação
  • Salvador C; Applied Electrochemistry Laboratory, Faculty of Experimental Sciences, University of Huelva, Av. 3 de Marzo s/n., 21007 Huelva, Spain.
  • Mesa MS; Department of Electronics Engineering, Computers and Automatic, ETSI, University of Huelva, Campus de La Rabida, 21810 Huelva, Spain.
  • Durán E; Department of Electronics Engineering, Computers and Automatic, ETSI, University of Huelva, Campus de La Rabida, 21810 Huelva, Spain.
  • Alvarez JL; Department of Information Technologies, ETSI, University of Huelva, Campus de La Rabida, 21810 Huelva, Spain.
  • Carbajo J; Applied Electrochemistry Laboratory, Faculty of Experimental Sciences, University of Huelva, Av. 3 de Marzo s/n., 21007 Huelva, Spain.
  • Mozo JD; Applied Electrochemistry Laboratory, Faculty of Experimental Sciences, University of Huelva, Av. 3 de Marzo s/n., 21007 Huelva, Spain.
Rev Sci Instrum ; 87(5): 055111, 2016 05.
Article em En | MEDLINE | ID: mdl-27250474
In this work, a new open hardware interface based on Arduino to read electromotive force (emf) from potentiometric detectors is presented. The interface has been fully designed with the open code philosophy and all documentation will be accessible on web. The paper describes a comprehensive project including the electronic design, the firmware loaded on Arduino, and the Java-coded graphical user interface to load data in a computer (PC or Mac) for processing. The prototype was tested by measuring the calibration curve of a detector. As detection element, an active poly(vinyl chloride)-based membrane was used, doped with cetyltrimethylammonium dodecylsulphate (CTA(+)-DS(-)). The experimental measures of emf indicate Nernstian behaviour with the CTA(+) content of test solutions, as it was described in the literature, proving the validity of the developed prototype. A comparative analysis of performance was made by using the same chemical detector but changing the measurement instrumentation.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies Idioma: En Revista: Rev Sci Instrum Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Espanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies Idioma: En Revista: Rev Sci Instrum Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Espanha