Quantification of myocardial blood flow by CZT-SPECT with motion correction and comparison with 15O-water PET.
Otaki, Yuka; Manabe, Osamu; Miller, Robert J H; Manrique, Alain; Nganoa, Catherine; Roth, Nathaniel; Berman, Daniel S; Germano, Guido; Slomka, Piotr J; Agostini, Denis.
Afiliação
  • Otaki Y; Departments of Imaging and Medicine, Cedars-Sinai Medical Center, Los Angeles, CA, USA.
  • Manabe O; Departments of Imaging and Medicine, Cedars-Sinai Medical Center, Los Angeles, CA, USA.
  • Miller RJH; Department of Nuclear Medicine, Hokkaido University of Graduate School of Medicine, Sapporo, Japan.
  • Manrique A; Departments of Imaging and Medicine, Cedars-Sinai Medical Center, Los Angeles, CA, USA.
  • Nganoa C; Department of Nuclear Medicine, CHU Cote de Nacre, Normandy University, Caen, France.
  • Roth N; Department of Nuclear Medicine, CHU Cote de Nacre, Normandy University, Caen, France.
  • Berman DS; Spectrum Dynamics Medical Ltd., Caesarea, Israel.
  • Germano G; Departments of Imaging and Medicine, Cedars-Sinai Medical Center, Los Angeles, CA, USA.
  • Slomka PJ; Departments of Imaging and Medicine, Cedars-Sinai Medical Center, Los Angeles, CA, USA.
  • Agostini D; Departments of Imaging and Medicine, Cedars-Sinai Medical Center, Los Angeles, CA, USA. slomkap@cshs.org.
J Nucl Cardiol ; 28(4): 1477-1486, 2021 Aug.
Article em En | MEDLINE | ID: mdl-31452085
ABSTRACT

BACKGROUND:

We compared quantification of MBF and myocardial flow reserve (MFR) with a 99mTc-sestamibi CZT-SPECT to 15O-water PET.

METHODS:

SPECT MBF for thirty patients in the WATERDAY study was re-analyzed by QPET software with motion correction and optimal placement of the arterial input function. 15O-water PET MBF was re-quantified using dedicated software. Inter-operator variability was assessed using repeatability coefficients (RPC).

RESULTS:

Significant correlations were observed between global (r = 0.91, P < 0.001) and regional MBF (r = 0.86, P < 0.001) with SPECT compared to PET. Global MBF (rest 0.95 vs 1.05 ml/min/g, P = 0.07; stress 2.62 vs 2.68 mL/min/g, P = 0.17) and MFR (2.65 vs 2.75, P = 0.86) were similar between SPECT and PET. Rest (0.81 vs 0.98 mL/min/g, P = 0.03) and stress MBF (1.98 vs 2.61 mL/min/g, P = 0.01) in right coronary artery (RCA) were lower with SPECT compared to PET. However, MFR in the RCA territory was similar (2.54 vs 2.77, P = 0.21). The SPECT-PET RPC for global MBFs and MFR were 0.95 mL/min/g and 0.94, with inter-observer RPC of 0.59 mL/min/g and 0.74, respectively.

CONCLUSIONS:

MBF and MFR derived from CZT-SPECT with motion correction and optimal placement of the arterial input function showed good agreement with 15O-water PET, as well as low inter-operator variability.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Telúrio / Zinco / Doença da Artéria Coronariana / Cádmio / Tomografia Computadorizada de Emissão de Fóton Único / Tomografia por Emissão de Pósitrons / Reserva Fracionada de Fluxo Miocárdico / Imagem de Perfusão do Miocárdio Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Nucl Cardiol Assunto da revista: CARDIOLOGIA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Telúrio / Zinco / Doença da Artéria Coronariana / Cádmio / Tomografia Computadorizada de Emissão de Fóton Único / Tomografia por Emissão de Pósitrons / Reserva Fracionada de Fluxo Miocárdico / Imagem de Perfusão do Miocárdio Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Nucl Cardiol Assunto da revista: CARDIOLOGIA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Estados Unidos