Ex vivo Comparison of 2 Disposable versus a State-of-the-Art Reusable Ureterorenoscope.
Bourdoumis, Andreas; El Howairis, Mohammed El Fatih; Stephen, Binu; Buchholz, Noor.
Afiliação
  • Bourdoumis A; U-merge Ltd. (Urology for Emerging Countries), London, United Kingdom.
  • El Howairis MEF; Department of Urology, Pennine Acute Hospitals NHS Trust, Northern Care Alliance, Manchester, United Kingdom.
  • Stephen B; U-merge Ltd. (Urology for Emerging Countries), London, United Kingdom.
  • Buchholz N; Department of Urology, Mediclinic City Hospital, Dubai, United Arab Emirates.
Urol Int ; 104(5-6): 437-444, 2020.
Article em En | MEDLINE | ID: mdl-31466078
INTRODUCTION: The success of upper urinary tract endoscopy on one hand and the high cost of purchase and maintenance of the scopes on the other hand have driven the development of disposable digital flexible ureterorenoscopy (fURS). MATERIALS AND METHODS: We performed an ex vivo comparison of 2 disposable digital fURS from different companies with a state-of-the-art reusable endoscope from another well-established scope manufacturer. The scopes were tested comparing parameters such as instrument handling, irrigation flow, image quality, and deflection. For the disposable scopes, the latter 2 were evaluated both, before and after vigorous stress testing. RESULTS: Expectedly, the most consistent finding among the 3 instruments was the marked decrease on irrigation flow following insertion of a working element, most notably a guidewire. Depth perception and ergonomic handling were comparable in all 3 instruments. Poststress testing revealed a significant loss of upward deflection in both disposable instruments, more so when an auxiliary instrument is inserted into the working channel. However, this did not appear to be the case in downward deflection. CONCLUSIONS: To our knowledge, this is the first study that used a vigorous stress testing protocol comparing disposable and reusable fURS. The newer generation disposable scope fared better than the older one. Results are encouraging and indicate that improvements are ongoing and progressing, bringing the quality of disposable scopes to the level that we are used to from our reusable high performance scopes while at the same time avoiding costly damage repairs. Continuous research and evidence-based interpretation of results in the absence of commercial bias are paramount to ensure this ongoing development.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Reutilização de Equipamento / Ureteroscópios / Equipamentos Descartáveis Idioma: En Revista: Urol Int Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Reino Unido

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Reutilização de Equipamento / Ureteroscópios / Equipamentos Descartáveis Idioma: En Revista: Urol Int Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Reino Unido