Your browser doesn't support javascript.
loading
Precision pulse capsulotomy complicated by radial tear of the anterior capsule: a proposed mechanism.
Safuri, Shadi; Duvdevan, Nitsan; Socea, Sergiu; Blumenthal, Eytan Z.
Afiliação
  • Safuri S; Department of Ophthalmology, Rambam Health Care Campus, Haifa, Israel. Electronic address: shadi.safuri@gmail.com.
  • Duvdevan N; Department of Ophthalmology, Rambam Health Care Campus, Haifa, Israel.
  • Socea S; Department of Ophthalmology, Rambam Health Care Campus, Haifa, Israel.
  • Blumenthal EZ; Department of Ophthalmology, Rambam Health Care Campus, Haifa, Israel; School of Medicine, Technion Israel Institute of Technology, Haifa, Israel.
J Cataract Refract Surg ; 45(11): 1680-1681, 2019 11.
Article em En | MEDLINE | ID: mdl-31706521
ABSTRACT
Zepto precision pulsed capsulotomy is an emerging technology aimed at providing a safer and more reproducible anterior capsulorhexis, with potential advantages in challenging cases. Initial reports suggest high safety, and thus far to our knowledge, no complications have been reported. Herein we report an unexpected complication. After the pulse delivery phase on a routine cataract case, a radial tear of the anterior capsule was observed. Upon careful review of the surgery video, an air bubble was noted, at the precise clock hour of the radial tear, trapped between the device wire and capsule. This air bubble presumably prevented the transfer of rapid phase transition at this site, interfering with capsule cleavage, and resulting in incomplete capsulotomy. Based on this observation, if a trapped air bubble is observed after the vacuum phase, we recommend applying more vacuum or disengaging and reattaching, before proceeding to the pulse delivery stage.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Facoemulsificação / Capsulorrexe / Microbolhas / Ruptura da Cápsula Anterior do Olho Tipo de estudo: Diagnostic_studies Limite: Aged / Humans / Male Idioma: En Revista: J Cataract Refract Surg Assunto da revista: OFTALMOLOGIA Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Facoemulsificação / Capsulorrexe / Microbolhas / Ruptura da Cápsula Anterior do Olho Tipo de estudo: Diagnostic_studies Limite: Aged / Humans / Male Idioma: En Revista: J Cataract Refract Surg Assunto da revista: OFTALMOLOGIA Ano de publicação: 2019 Tipo de documento: Article