Lactate measurement by neurochemical profiling in the dorsolateral prefrontal cortex at 7T: accuracy, precision, and relaxation times.
Dehghani, Masoumeh; Do, Kim Q; Magistretti, Pierre; Xin, Lijing.
Afiliação
  • Dehghani M; Center for Psychiatric Neuroscience (CNP), Department of Psychiatry, Lausanne University Hospital-CHUV, Prilly-Lausanne, Switzerland.
  • Do KQ; Center for Psychiatric Neuroscience (CNP), Department of Psychiatry, Lausanne University Hospital-CHUV, Prilly-Lausanne, Switzerland.
  • Magistretti P; Center for Psychiatric Neuroscience (CNP), Department of Psychiatry, Lausanne University Hospital-CHUV, Prilly-Lausanne, Switzerland.
  • Xin L; BESE Division, King Abdullah University of Science and Technology, Thuwal, Saudi Arabia.
Magn Reson Med ; 83(6): 1895-1908, 2020 06.
Article em En | MEDLINE | ID: mdl-31729080
ABSTRACT

PURPOSE:

This assesses the potential of measuring lactate in the human brain using three non-editing MRS methods at 7T and compares the accuracy and precision of the methods.

METHODS:

1 H MRS data were measured in the right dorsolateral prefrontal cortex using a semi-adiabatic spin-echo full-intensity acquired localized sequence with three different protocols (I) TE = 16 ms, (II) TE = 110 ms, and (III) TE = 16 ms, TI = 300 ms. T1 and T2 relaxation times of lactate were also measured. Simulated spectra were generated for three protocols with known concentrations, using a range of spectral linewidths and SNRs to assess the effect of data quality on the measurement precision and accuracy.

RESULTS:

Lactate was quantified in all three protocols with mean Cramér-Rao lower bound of 8% (I), 13% (II), and 7% (III). The T1 and T2 relaxation times of lactate were 1.9 ± 0.2 s and 94 ± 13 ms, respectively. Simulations predicted a spectral linewidth-associated underestimation of lactate measurement. Simulations, phantom and in vivo results showed that protocol II was most affected by this underestimation. In addition, the estimation error was insensitive to a broad range of spectral linewidth with protocol I. Within-session coefficient of variances of lactate were 6.1 ± 7.9% (I), 22.3 ± 12.3% (II), and 5.1 ± 5.4% (III), respectively.

CONCLUSION:

We conclude that protocols I and III have the potential to measure lactate at 7T with good reproducibility, whereas the measurement accuracy and precision depend on spectral linewidth and SNR, respectively. Moreover, simulation is valuable for the optimization of measurement protocols in future study design and the correction for measurement bias.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Encéfalo / Córtex Pré-Frontal Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Suíça

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Encéfalo / Córtex Pré-Frontal Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Suíça