Using force data to self-pace an instrumented treadmill and measure self-selected walking speed.
Song, Seungmoon; Choi, Hojung; Collins, Steven H.
Afiliação
  • Song S; Department of Mechanical Engineering, Stanford University, Stanford, CA, USA. smsong@stanford.edu.
  • Choi H; Department of Mechanical Engineering, Stanford University, Stanford, CA, USA.
  • Collins SH; Department of Mechanical Engineering, Stanford University, Stanford, CA, USA.
J Neuroeng Rehabil ; 17(1): 68, 2020 06 03.
Article em En | MEDLINE | ID: mdl-32493426
ABSTRACT

BACKGROUND:

Self-selected speed is an important functional index of walking. A self-pacing controller that reliably matches walking speed without additional hardware can be useful for measuring self-selected speed in a treadmill-based laboratory.

METHODS:

We adapted a previously proposed self-pacing controller for force-instrumented treadmills and validated its use for measuring self-selected speeds. We first evaluated the controller's estimation of subject speed and position from the force-plates by comparing it to those from motion capture data. We then compared five tests of self-selected speed. Ten healthy adults completed a standard 10-meter walk test, a 150-meter walk test, a commonly used manual treadmill speed selection test, a two-minute self-paced treadmill test, and a 150-meter self-paced treadmill test. In each case, subjects were instructed to walk at or select their comfortable speed. We also assessed the time taken for a trial and a survey on comfort and ease of choosing a speed in all the tests.

RESULTS:

The self-pacing algorithm estimated subject speed and position accurately, with root mean square differences compared to motion capture of 0.023 m s -1 and 0.014 m, respectively. Self-selected speeds from both self-paced treadmill tests correlated well with those from the 10-meter walk test (R>0.93,p<1×10-13). Subjects walked slower on average in the self-paced treadmill tests (1.23±0.27 ms-1) than in the 10-meter walk test (1.32±0.18 ms-1) but the speed differences within subjects were consistent. These correlations and walking speeds are comparable to those from the manual treadmill speed selection test (R=0.89,p=3×10-11;1.18±0.24 ms-1). Comfort and ease of speed selection were similar in the self-paced tests and the manual speed selection test, but the self-paced tests required only about a third of the time to complete. Our results demonstrate that these self-paced treadmill tests can be a strong alternative to the commonly used manual treadmill speed selection test.

CONCLUSIONS:

The self-paced force-instrumented treadmill well adapts to subject walking speed and reliably measures self-selected walking speeds. We provide the self-pacing software to facilitate use by gait researchers and clinicians.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Algoritmos / Teste de Esforço / Velocidade de Caminhada Tipo de estudo: Guideline Limite: Adult / Female / Humans / Male Idioma: En Revista: J Neuroeng Rehabil Assunto da revista: ENGENHARIA BIOMEDICA / NEUROLOGIA / REABILITACAO Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Algoritmos / Teste de Esforço / Velocidade de Caminhada Tipo de estudo: Guideline Limite: Adult / Female / Humans / Male Idioma: En Revista: J Neuroeng Rehabil Assunto da revista: ENGENHARIA BIOMEDICA / NEUROLOGIA / REABILITACAO Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Estados Unidos