Your browser doesn't support javascript.
loading
Proton therapy range verification method via delayed γ-ray spectroscopy of a molybdenum tumour marker.
Burbadge, C; Kasanda, E; Bildstein, V; Dublin, G; Olaizola, B; Höhr, C; Mücher, D.
Afiliação
  • Burbadge C; Department of Physics, University of Guelph, 50 Stone Rd E, Guelph, Ontario, N1G 2W1, Canada.
  • Kasanda E; Department of Physics, University of Guelph, 50 Stone Rd E, Guelph, Ontario, N1G 2W1, Canada.
  • Bildstein V; Department of Physics, University of Guelph, 50 Stone Rd E, Guelph, Ontario, N1G 2W1, Canada.
  • Dublin G; Department of Physics, University of Guelph, 50 Stone Rd E, Guelph, Ontario, N1G 2W1, Canada.
  • Olaizola B; TRIUMF, 4004 Wesbrook Mall, Vancouver, British Columbia, V6T 2A3, Canada.
  • Höhr C; TRIUMF, 4004 Wesbrook Mall, Vancouver, British Columbia, V6T 2A3, Canada.
  • Mücher D; Department of Physics and Astronomy, University of Victoria, 3800 Finnerty Road, Victoria, British Columbia, V8P 5C2, Canada.
Phys Med Biol ; 66(2): 025005, 2021 01 26.
Article em En | MEDLINE | ID: mdl-32998122
ABSTRACT
In this work, a new method of range verification for proton therapy (PT) is experimentally demonstrated for the first time. If a metal marker is implanted near the tumour site, its response to proton activation will result in the emission of characteristic γ rays. The relative intensity of γ rays originating from competing fusion-evaporation reaction channels provides a unique signature of the average proton energy at the marker, and by extension the beam's range, in vivo and in real time. The clinical feasibility of this method was investigated at the PT facility at TRIUMF with a proof-of-principle experiment which irradiated a naturally-abundant molybdenum foil at various proton beam energies. Delayed characteristic γ rays were measured with two Compton-shielded LaBr3 scintillators. The technique was successfully demonstrated by relating the relative intensity of two γ-ray peaks to the energy of the beam at the Mo target, opening the door to future clinical applications where the range of the beam can be verified in real time.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Análise Espectral / Terapia com Prótons / Raios gama / Molibdênio Tipo de estudo: Health_economic_evaluation Limite: Humans Idioma: En Revista: Phys Med Biol Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Canadá

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Análise Espectral / Terapia com Prótons / Raios gama / Molibdênio Tipo de estudo: Health_economic_evaluation Limite: Humans Idioma: En Revista: Phys Med Biol Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Canadá