Your browser doesn't support javascript.
loading
A Two-Electrode, Double-Pulsed Sensor Readout Circuit for Cardiac Troponin I Measurement.
IEEE Trans Biomed Circuits Syst ; 14(6): 1362-1370, 2020 12.
Article em En | MEDLINE | ID: mdl-33035161
This paper presents a pulse-stimulus sensor readout circuit for use in cardiovascular disease examinations. The sensor is based on a gold nanoparticle plate with an antibody post-modification. The proposed system utilizes gated pulses to detect the biomarker Cardiac Troponin I in an ionic solution. The characteristic of the electrostatic double-layer capacitor generated by the analyte is related to the concentration of Cardiac Troponin I in the solvent. After sensing by the transistor, a current-to-frequency converter (I-to-F) and delay-line-based time-to-digital converter (TDC) convert the information into a series of digital codes for further analysis. The design is fabricated in a 0.18-µm standard CMOS process. The chip occupies an area of 0.92 mm2 and consumes 125 µW. In the measurements, the proposed circuit achieved a 1.77 Hz/pg-mL sensitivity and 72.43 dB dynamic range.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Técnicas Biossensoriais / Troponina I Limite: Humans Idioma: En Revista: IEEE Trans Biomed Circuits Syst Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Técnicas Biossensoriais / Troponina I Limite: Humans Idioma: En Revista: IEEE Trans Biomed Circuits Syst Ano de publicação: 2020 Tipo de documento: Article