Swept Source Optical Coherence Tomography Compared to Ultrasound and Biomicroscopy for Diagnosis of Posterior Vitreous Detachment.
Wang, Marlene D; Truong, Cecile; Mammo, Zaid; Hussnain, Syed Amal; Chen, Royce W S.
Afiliação
  • Wang MD; Department of Ophthalmology, Edward S. Harkness Eye Institute, Columbia University Irving Medical Center, New York, NY, 10032, USA.
  • Truong C; Department of Ophthalmology, Edward S. Harkness Eye Institute, Columbia University Irving Medical Center, New York, NY, 10032, USA.
  • Mammo Z; Department of Ophthalmology, Edward S. Harkness Eye Institute, Columbia University Irving Medical Center, New York, NY, 10032, USA.
  • Hussnain SA; Department of Ophthalmology, Edward S. Harkness Eye Institute, Columbia University Irving Medical Center, New York, NY, 10032, USA.
  • Chen RWS; Department of Ophthalmology, Edward S. Harkness Eye Institute, Columbia University Irving Medical Center, New York, NY, 10032, USA.
Clin Ophthalmol ; 15: 507-512, 2021.
Article em En | MEDLINE | ID: mdl-33603328
ABSTRACT

BACKGROUND:

Biomicroscopy, B-scan ultrasound imaging, and SD-OCT are all modalities used to characterize a posterior vitreous detachment (PVD). Our objective is to assess the precision of the diagnosis of PVD by SS-OCT.

METHODS:

This prospective observational study examines ninety-five eyes of forty-nine patients with biomicroscopy, B-scan ultrasound, and SS-OCT for the presence or absence of a complete PVD. All SS-OCT images were reviewed by two retina specialists (RWSC, ZM). All three diagnostic methods were evaluated for agreement by Cohen's kappa statistic.

RESULTS:

The inter-rater reliability between retina specialists reading the SS-OCT images was 97.9% (κ = 0.957). Agreement on PVD status between SS-OCT and biomicroscopy was 85.3% (κ = 0.711). Agreement between SS-OCT and B-scan ultrasound was 83.2% (κ = 0.667). Agreement between B-scan ultrasound and biomicroscopy was 87.4% (κ = 0.743).

CONCLUSION:

For the diagnosis of complete PVD, SS-OCT allows for high accuracy and agreement between graders.