Treatment of air leakage using the VIO soft coagulation system: a mouse pulmonary air leak model.
Takahashi, Yuki; Saito, Atsushi; Sakuma, Yuji; Tada, Makoto; Maki, Ryunosuke; Takahashi, Motoko; Watanabe, Atsushi.
Afiliação
  • Takahashi Y; Department of Thoracic Surgery, School of Medicine and Hospital, Sapporo Medical University, South 1, West 16, Chuo-ku, Sapporo, Hokkaido, 060-8543, Japan.
  • Saito A; Department of Respiratory Medicine and Allergology, School of Medicine and Hospital, Sapporo Medical University, South 1, West 16, Chuo-ku, Sapporo, Hokkaido, 060-8543, Japan.
  • Sakuma Y; Department of Molecular Medicine, Research Institute for Frontier Medicine, Sapporo Medical University School of Medicine, South 1, West 17, Chuo-ku, Sapporo, Hokkaido, 060-8556, Japan.
  • Tada M; Department of Thoracic Surgery, School of Medicine and Hospital, Sapporo Medical University, South 1, West 16, Chuo-ku, Sapporo, Hokkaido, 060-8543, Japan.
  • Maki R; Department of Thoracic Surgery, School of Medicine and Hospital, Sapporo Medical University, South 1, West 16, Chuo-ku, Sapporo, Hokkaido, 060-8543, Japan.
  • Takahashi M; Department of Biochemistry, Sapporo Medical University School of Medicine, South 1, West 17, Chuo-ku, Sapporo, Hokkaido, 060-8556, Japan.
  • Watanabe A; Department of Thoracic Surgery, School of Medicine and Hospital, Sapporo Medical University, South 1, West 16, Chuo-ku, Sapporo, Hokkaido, 060-8543, Japan. atsushiw@sapmed.ac.jp.
Surg Today ; 51(9): 1521-1529, 2021 Sep.
Article em En | MEDLINE | ID: mdl-33743052
ABSTRACT

PURPOSE:

We aimed to compare the efficacy of the VIO soft coagulation system (VSCS) for the treatment of air leaks by sealing with fibrin glue, and also assess the histological alterations that occur after soft coagulation.

METHODS:

A mouse pulmonary air leak model was designed. The pulmonary fistula was subsequently coagulated with the VSCS or sealed with fibrin glue with polyglycolic acid (PGA) sheets. The burst pressure at air leak recurrence was measured in each group, and the results were compared. We also evaluated the histological alterations in the mouse pulmonary air leak model after soft coagulation with the VSCS.

RESULTS:

The burst pressure in the soft coagulation group (80 W/Effect 5) (median 42.8; range 35.4-53.8 cmH2O) was similar to that in the fibrin glue group (median 41.5; range 34.6-43.9 cmH2O) (p = 0.21). Histological examinations revealed that the visceral pleura remained torn, the structure of the pulmonary alveolus was maintained, and the coagulated fistula was covered with a fibrin membrane in the soft coagulation group.

CONCLUSIONS:

The pressure resistance following soft coagulation was equivalent to that after sealing using fibrin glue with PGA sheets. The air leaks were likely controlled by covering the fistula with a fibrin membrane after soft coagulation with the VSCS.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Pneumonectomia / Complicações Pós-Operatórias / Ar / Eletrocoagulação / Fístula Anastomótica Tipo de estudo: Etiology_studies Limite: Animals Idioma: En Revista: Surg Today Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Japão

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Pneumonectomia / Complicações Pós-Operatórias / Ar / Eletrocoagulação / Fístula Anastomótica Tipo de estudo: Etiology_studies Limite: Animals Idioma: En Revista: Surg Today Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Japão