Your browser doesn't support javascript.
loading
Automated treatment planning of prostate stereotactic body radiotherapy with focal boosting on a fast-rotating O-ring linac: Plan quality comparison with C-arm linacs.
De Roover, Robin; Crijns, Wouter; Poels, Kenneth; Dewit, Bertrand; Draulans, Cédric; Haustermans, Karin; Depuydt, Tom.
Afiliação
  • De Roover R; Department of Radiation Oncology, University Hospitals Leuven, Leuven, Belgium.
  • Crijns W; Department of Oncology, KU Leuven, Leuven, Belgium.
  • Poels K; Department of Radiation Oncology, University Hospitals Leuven, Leuven, Belgium.
  • Dewit B; Department of Oncology, KU Leuven, Leuven, Belgium.
  • Draulans C; Department of Radiation Oncology, University Hospitals Leuven, Leuven, Belgium.
  • Haustermans K; Department of Oncology, KU Leuven, Leuven, Belgium.
  • Depuydt T; Department of Radiation Oncology, University Hospitals Leuven, Leuven, Belgium.
J Appl Clin Med Phys ; 22(9): 59-72, 2021 Sep.
Article em En | MEDLINE | ID: mdl-34318996
PURPOSE: The integration of auto-segmentation and automated treatment planning methods on a fast-rotating O-ring linac may improve the time efficiency of online adaptive radiotherapy workflows. This study investigates whether automated treatment planning of prostate SBRT with focal boosting on the O-ring linac could generate plans that are of similar quality as those obtained through manual planning on clinical C-arm linacs. METHODS: For 20 men with prostate cancer, reference treatment plans were generated on a TrueBeam STx C-arm linac with HD120 MLC and a TrueBeam C-arm linac with Millennium 120 MLC using 6 MV flattened dual arc VMAT. Manual planning on the Halcyon fast-rotating O-ring linac was performed using 6 MV FFF dual arc VMAT (HA2-DL10) and triple arc VMAT (HA3-DL10) to investigate the performance of the dual-layer MLC system. Automated planning was performed for triple arc VMAT on the Halcyon linac (ET3-DL10) using the automated planning algorithms of Ethos Treatment Planning. The prescribed dose was 35 Gy to the prostate and 30 Gy to the seminal vesicles in five fractions. The iso-toxic focal boost to the intraprostatic tumor nodule(s) was aimed to receive up to 50 Gy. Plan deliverability was verified using portal image dosimetry measurements. RESULTS: Compared to the C-arm linacs, ET3-DL10 shows increased seminal vesicles PTV coverage (D99% ) and reduced high-dose spillage to the bladder (V37Gy ) and urethra (D0.035cc ) but this came at the cost of increased high-dose spillage to the rectum (V38Gy ) and a higher intermediate dose spillage (D2cm). No statistically significant differences were found when benchmarking HA2-DL10 and HA3-DL10 with the C-arm linacs. All plans passed the patient-specific QA tolerance limit. CONCLUSIONS: Automated planning of prostate SBRT with focal boosting on the fast-rotating O-ring linac is feasible and achieves similar plan quality as those obtained on clinical C-arm linacs using manual planning.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiocirurgia / Radioterapia de Intensidade Modulada Tipo de estudo: Guideline Limite: Humans / Male Idioma: En Revista: J Appl Clin Med Phys Assunto da revista: BIOFISICA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Bélgica

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiocirurgia / Radioterapia de Intensidade Modulada Tipo de estudo: Guideline Limite: Humans / Male Idioma: En Revista: J Appl Clin Med Phys Assunto da revista: BIOFISICA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Bélgica