What Are the Risk Factors for Mechanical Failure and Loosening of a Transfemoral Osseointegrated Implant System in Patients with a Lower-limb Amputation?
Mohamed, Jamal; Reetz, David; van de Meent, Henk; Schreuder, Hendrik; Frölke, Jan Paul; Leijendekkers, Ruud.
Afiliação
  • Mohamed J; Department of Surgery, Radboud University Medical Center, Nijmegen, the Netherlands.
  • Reetz D; Department of Surgery, Radboud University Medical Center, Nijmegen, the Netherlands.
  • van de Meent H; Orthopedic Research Laboratory, Radboud Institute for Health Sciences, Radboud University Medical Center, Nijmegen, the Netherlands.
  • Schreuder H; Department of Rehabilitation, Radboud University Medical Center, Nijmegen, the Netherlands.
  • Frölke JP; Department of Orthopedics, Radboud University Medical Center, Nijmegen, the Netherlands.
  • Leijendekkers R; Department of Surgery, Radboud University Medical Center, Nijmegen, the Netherlands.
Clin Orthop Relat Res ; 480(4): 722-731, 2022 04 01.
Article em En | MEDLINE | ID: mdl-34889879
ABSTRACT

BACKGROUND:

Septic loosening and stem breakage due to metal fatigue is a rare but well-known cause of orthopaedic implant failure. This may also affect the components of the osseointegrated implant system for individuals with transfemoral amputation who subsequently undergo revision. Identifying risk factors is important to minimize the frequency of revision surgery after implant breakage. QUESTIONS/

PURPOSES:

(1) What proportion of patients who received an osseointegrated implant after transfemoral amputation underwent revision surgery, and what were the causes of those revisions? (2) What factors were associated with revision surgery when stratified by the location of the mechanical failure and (septic) loosening (intramedullary stem versus dual cone adapter)?

METHODS:

Between May 2009 and July 2015, we treated 72 patients with an osseointegrated implant. Inclusion criteria were a minimum follow-up of 5-years and a standard press-fit cobalt-chromium-molybdenum (CoCrMb) transfemoral osseointegrated implant. Based on that, 83% (60 of 72) of patients were eligible; a further 3% (2 of 60) were excluded because of no received informed consent (n = 1) and loss to follow-up (n = 1). Eventually, we included 81% (58 of 72) of patients for analysis in this retrospective, comparative study. We compared patient characteristics (gender, age, and BMI), implant details (diameter of the intramedullary stem, length of the dual cone, and implant survival time), and event characteristics (infectious complications and distal bone resorption). The data were retrieved from our electronic patient file and from our cloud-based database and analyzed by individuals not involved in patient care. Failures were categorized as (1) mechanical failures, defined as breakage of the intramedullary stem or dual-cone adapter, or (2) (septic) loosening of the osseointegrated implant.

RESULTS:

Thirty-four percent (20 of 58) of patients had revision surgery. In 12% (7 of 58) of patients, the reason for revision was due to intramedullary stem failures (six breakages, one septic loosening), and in 22% (13 of 58) of patients it was due to dual-cone adaptor failure (10 weak-point breakages and four distal taper breakages; one patient broke both the weak-point and the dual-cone adapter). Smaller median stem diameter (failure 15 mm [interquartile range 1.3], nonfailure 17 mm [IQR 2.0], difference of medians 2 mm; p < 0.01) and higher median number of infectious events (failure 6 [IQR 11], nonfailure 1 [IQR 3.0], difference of medians -5; p < 0.01) were associated with revision intramedullary stem surgery. No risk factors could be identified for broken dual-cone adapters.

CONCLUSION:

Possible risk factors for system failure of this osteointegration implant include small stem diameter and high number of infectious events. We did not find factors associated with dual-cone adapter weak-point failure and distal taper failure, most likely because of the small sample size. When treating a person with a lower-limb amputation with a CoCrMb osseointegrated implant, we recommend avoiding a small stem diameter. Further research with longer follow-up is needed to study the success of revised patients. LEVEL OF EVIDENCE Level III, therapeutic study.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Prótese Ancorada no Osso Tipo de estudo: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: Clin Orthop Relat Res Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Holanda

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Prótese Ancorada no Osso Tipo de estudo: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: Clin Orthop Relat Res Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Holanda