Quantitative 177Lu SPECT/CT imaging for personalized dosimetry using a ring-shaped CZT-based camera.
Danieli, Rachele; Stella, Martina; Leube, Julian; Tran-Gia, Johannes; Marin, Clementine; Uribe, Carlos F; Vanderlinden, Bruno; Reynaert, Nick; Flamen, Patrick; Levillain, Hugo.
Afiliação
  • Danieli R; Department of Medical Physics, Institut Jules Bordet, Hôpital Universitaire de Bruxelles (H.U.B), Université Libre de Bruxelles (ULB), Brussels, Belgium. rachele.danieli@hubruxelles.be.
  • Stella M; Radiophysics and MRI Physics Laboratory, Université Libre de Bruxelles (ULB), Brussels, Belgium. rachele.danieli@hubruxelles.be.
  • Leube J; Radiophysics and MRI Physics Laboratory, Université Libre de Bruxelles (ULB), Brussels, Belgium.
  • Tran-Gia J; GE HealthCare, Diegem, Belgium.
  • Marin C; Department of Nuclear Medicine, University Hospital Würzburg, Würzburg, Germany.
  • Uribe CF; Department of Nuclear Medicine, University Hospital Würzburg, Würzburg, Germany.
  • Vanderlinden B; Department of Medical Physics, Institut Jules Bordet, Hôpital Universitaire de Bruxelles (H.U.B), Université Libre de Bruxelles (ULB), Brussels, Belgium.
  • Reynaert N; Functional Imaging, BC Cancer, Vancouver, BC, Canada.
  • Flamen P; Department of Medical Physics, Institut Jules Bordet, Hôpital Universitaire de Bruxelles (H.U.B), Université Libre de Bruxelles (ULB), Brussels, Belgium.
  • Levillain H; Radiophysics and MRI Physics Laboratory, Université Libre de Bruxelles (ULB), Brussels, Belgium.
EJNMMI Phys ; 10(1): 64, 2023 Oct 18.
Article em En | MEDLINE | ID: mdl-37853247
ABSTRACT

BACKGROUND:

Dosimetry after radiopharmaceutical therapy with 177Lu (177Lu-RPT) relies on quantitative SPECT/CT imaging, for which suitable reconstruction protocols are required. In this study, we characterized for the first time the quantitative performance of a ring-shaped CZT-based camera using two different reconstruction algorithms an ordered subset expectation maximization (OSEM) and a block sequential regularized expectation maximization (BSREM) combined with noise reduction regularization. This study lays the foundations for the definition of a reconstruction protocol enabling accurate dosimetry for patients treated with 177Lu-RPT.

METHODS:

A series of 177Lu-filled phantoms were acquired on a StarGuide™ (GE HealthCare), with energy and scatter windows centred at 208 (± 6%) keV and 185 (± 5%) keV, respectively. Images were reconstructed with the manufacturer implementations of OSEM (GE-OSEM) and BSREM (Q.Clear) algorithms, and various combinations of iterations and subsets. Additionally, the manufacturer-recommended Q.Clear-based reconstruction protocol was evaluated. Quantification accuracy, measured as the difference between the SPECT-based and the radionuclide calibrator-based activity, and noise were evaluated in a large cylinder. Recovery coefficients (RCs) and spatial resolution were assessed in a NEMA IEC phantom with sphere inserts. The reconstruction protocols considered suitable for clinical applications were tested on a cohort of patients treated with [177Lu]Lu-PSMA-I&T.

RESULTS:

The accuracy of the activity from the cylinder, although affected by septal penetration, was < 10% for all reconstructions. Both algorithms featured improved spatial resolution and higher RCs with increasing updates at the cost of noise build-up, but Q.Clear outperformed GE-OSEM in reducing noise accumulation. When the reconstruction parameters were carefully selected, similar values for noise (~0.15), spatial resolution (~1 cm) and RCs were found, irrespective of the reconstruction algorithm. Analogue results were found in patients.

CONCLUSIONS:

Accurate activity quantification is possible when imaging 177Lu with StarGuide™. However, the impact of septal penetration requires further investigations. GE-OSEM is a valid alternative to the recommended Q.Clear reconstruction algorithm, featuring comparable performances assessed on phantoms and patients.