Automatic adjustment of oxygen concentration during high-flow nasal cannula treatment using a targeted SpO2 feedback system.
Seo, Woo Jung; Kim, Eun Young; Seo, Ga Jin; Suh, Hee Jung; Huh, Jin Won; Hong, Sang-Bum; Koh, Younsuck; Lim, Chae-Man.
Afiliação
  • Seo WJ; Division of Pulmonary and Critical Care Medicine, Department of Internal Medicine, Ilsan Paik Hospital, Inje University College of Medicine, Goyang, Republic of Korea.
  • Kim EY; Intensive Care Nursing Team, Asan Medical Center, Seoul, Republic of Korea.
  • Seo GJ; Intensive Care Nursing Team, Asan Medical Center, Seoul, Republic of Korea.
  • Suh HJ; Intensive Care Nursing Team, Asan Medical Center, Seoul, Republic of Korea.
  • Huh JW; Department of Pulmonary and Critical Care Medicine, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Republic of Korea.
  • Hong SB; Department of Pulmonary and Critical Care Medicine, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Republic of Korea.
  • Koh Y; Department of Pulmonary and Critical Care Medicine, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Republic of Korea.
  • Lim CM; Department of Pulmonary and Critical Care Medicine, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Republic of Korea.
Nurs Crit Care ; 2024 Feb 06.
Article em En | MEDLINE | ID: mdl-38320812
ABSTRACT

AIM:

To test whether targeted SpO2 feedback (TSF), an automatic control system for fraction of inspired oxygen (FiO2), achieves more time in the optimal SpO2 range and/or reduces the frequency of manual adjustments to administered FiO2 compared with conventional manual titration in patients with hypoxia on high-flow nasal cannula (HFNC) therapy. STUDY

DESIGN:

Twenty-two patients were recruited from two hospitals. For each, two sessions of manual mode and two sessions of TSF were applied in a random order, each session lasting 2 h. The target SpO2 on TSF was 95%. Oxygen monitoring levels were classified into four SpO2 ranges hypoxia (≤ 89%), borderline (90%-93%), optimal (94%-96%) and hyperoxia (≥ 97%). The two modes were compared based on the proportion of time spent in each SpO2 range and the number of manual FiO2 adjustments.

RESULTS:

The proportion of time in the optimal SpO2 range was 20.5% under manual titration mode and 65.4% under TSF (p < .01). The proportions of time in the hypoxia range were 1.1% and 0.4%, respectively (p = .31), in the borderline range 4.7% and 3.5%, respectively (p = .54), and in the hyperoxia range 73.7% and 30.7%, respectively (p < .01). There were statistical differences only in the optimal and hyperoxia SpO2 ranges. During the 8 h, the frequency of manual FiO2 adjustment was 0.7 times for the manual mode and 0.2 times for TSF, showing no statistically significant difference (p = 0.076).

CONCLUSION:

Compared with manual titration, TSF achieved greater time of the optimal SpO2 and less time of hyperoxia during HFNC. The frequency of manual adjustments on TSF tended to be less than on manual titration mode. RELEVANCE TO CLINICAL PRACTICE Automatic closed-loop algorithm FiO2 monitoring systems can achieve better oxygen treatments than conventional monitoring and may reduce nurse workloads. In the era of pandemic respiratory diseases, this system can also facilitate contactless SpO2 monitoring during HFNC therapy.