Bicuspid valve CT registry of balloon-expandable TAVR: BETTER TAVR registry.
Catheter Cardiovasc Interv
; 104(1): 105-114, 2024 Jul.
Article
em En
| MEDLINE
| ID: mdl-38819623
ABSTRACT
BACKGROUND:
The anatomic substrate of bicuspid valves may lead to suboptimal TAVR stent expansion and geometry.AIM:
We evaluated determinants of stent geometry in bicuspid valves treated with Sapien transcatheter aortic valve replacement (TAVR) valves.METHODS:
A multicenter retrospective registry of patients (February 2019 to August 2022) who underwent post-TAVR computed tomography to determine stent area (vs. nominal valve area) and stent ellipticity (maximum diameter/minimum diameter). Predictors of relative stent expansion (minimum area/average of inflow + outflow area) and stent ellipticity were evaluated in a multivariable regression model, including valve calcium volume (indexed by annular area), presence of raphe calcium, sinus diameters indexed by area-derived annular diameter, and performance of pre-dilation and post-dilation.RESULTS:
The registry enrolled 101 patients from four centers. The minimum stent area (vs. nominal area) was 88.1%, and the maximum ellipticity was 1.10, with both observed near the midframe of the valve in all cases. Relative stent expansion ≥90% was observed in 64/101 patients. The only significant predictor of relative stent expansion ≥90% was the performance of post-dilation (OR 4.79, p = 0.018). Relative stent expansion ≥90% was seen in 86% of patients with post-dilation compared to 57% without (p < 0.001). The stent ellipticity ≥1.1 was observed in 47/101 patients. The significant predictors of stent ellipticity ≥1.1 were the indexed maximum sinus diameter (OR 0.582, p = 0.021) and indexed intercommisural diameter at 4 mm (OR 2.42, p = 0.001). Stent expansion has a weak negative correlation with post-TAVR mean gradient (r = -0.324, p < 0.001).CONCLUSION:
Relative stent expansion ≥90% was associated with the performance of post-dilation, and stent ellipticity ≥1.1 was associated with indexed intercommisural diameter and indexed maximum sinus diameter. Further studies to determine optimal deployment strategies in bicuspid valves are needed.Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Valva Aórtica
/
Estenose da Valva Aórtica
/
Desenho de Prótese
/
Próteses Valvulares Cardíacas
/
Sistema de Registros
/
Substituição da Valva Aórtica Transcateter
Limite:
Aged
/
Aged80
/
Female
/
Humans
/
Male
País/Região como assunto:
America do norte
Idioma:
En
Revista:
Catheter Cardiovasc Interv
Assunto da revista:
CARDIOLOGIA
Ano de publicação:
2024
Tipo de documento:
Article
País de afiliação:
Canadá