Your browser doesn't support javascript.
loading
Assessment of the Beckman creatinine analyzer 2.
Clin Biochem ; 12(2): 46-9, 1979 Apr.
Article em En | MEDLINE | ID: mdl-445793
ABSTRACT
An evaluation of the Beckman Creatinine Analyzer 2 has been carried out. Good day-to-day and within-day precision was obtained. Analyses of quality control materials, studies of linearity, and comparison of the results obtained for samples from patients with the results obtained by a continuous-flow method evidenced good accuracy. Aceto-acetate in serum caused significant positive interference. The desirable performance characteristics and practicability make this Analyzer attractive to the clinical chemist, particularly for paediatric and emergency laboratory use.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Creatinina Limite: Humans Idioma: En Revista: Clin Biochem Ano de publicação: 1979 Tipo de documento: Article
Buscar no Google