1.
Bioanalysis ; 15(6): 331-341, 2023 Mar.
Artículo en Inglés | MEDLINE | ID: mdl-36961376

RESUMEN

Aim: To develop a nondestructive method for the estimation of hematocrit (HCT) in dried blood spots (DBSs). Materials & methods: Standards and controls were created (HCT range: 0.20-0.50 l/l) and DBS scanned using a flatbed scanner. Gray values and pixel areas were analyzed with open-source software to estimate HCT and volume, respectively. HCT obtained in whole blood using hematological analyzer was compared with DBS scanner method (n = 50). Results: Between-run precision was 4.7-10.2% and between-run accuracy was 89.6-102.1%. In the hematological instrument comparison, 96% of the patient sample results were within ±15%. Conclusion: The nondestructive method can be used to exclude patient DBS samples with extreme HCT levels from further analysis and avoid bias on measured concentration.


Asunto(s)
Pruebas con Sangre Seca , Humanos , Pruebas con Sangre Seca/métodos , Hematócrito