1.
Radiat Prot Dosimetry ; 189(3): 354-361, 2020 Jul 17.
Artículo en Inglés | MEDLINE | ID: mdl-32342104

RESUMEN

In the absence of information on radiation doses in mammography in the Gulf countries, this study was designed to assess patient dose in terms of entrance surface air kerma and average glandular dose (AGD) in three mammography units in Qatar that covers 21% of all mammography systems in the country. The study of 150 patients involving 600 projections indicated that the average value of AGD in patients was 2.2 mGy for cranio-caudal and 2.5 mGy for mediolateral-oblique views, respectively. Dose assessment was also performed for polymethyl methacrylate phantoms of thicknesses, ranging from 20 to 80 mm. Comparing the patient dose values with several other publications in literature for full-field digital mammography, our values are typically higher, which can be likely attributed to the larger compressed breast thickness.


Asunto(s)
Mamografía , Intensificación de Imagen Radiográfica , Mama/diagnóstico por imagen , Humanos , Fantasmas de Imagen , Qatar , Dosis de Radiación